Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning

This work considers a previously overlooked uncertainty present in film dosimetry which results from moderate curvature of films during the scanning process. Small film samples are particularly susceptible to film curling which may be undetected or deemed insignificant. In this study, we consider te...

Descripción completa

Detalles Bibliográficos
Autores principales: Palmer, Antony L., Bradley, David A., Nisbet, Andrew
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690100/
https://www.ncbi.nlm.nih.gov/pubmed/26103181
http://dx.doi.org/10.1120/jacmp.v16i2.5141
_version_ 1783279529642426368
author Palmer, Antony L.
Bradley, David A.
Nisbet, Andrew
author_facet Palmer, Antony L.
Bradley, David A.
Nisbet, Andrew
author_sort Palmer, Antony L.
collection PubMed
description This work considers a previously overlooked uncertainty present in film dosimetry which results from moderate curvature of films during the scanning process. Small film samples are particularly susceptible to film curling which may be undetected or deemed insignificant. In this study, we consider test cases with controlled induced curvature of film and with film raised horizontally above the scanner plate. We also evaluate the difference in scans of a film irradiated with a typical brachytherapy dose distribution with the film naturally curved and with the film held flat on the scanner. Typical naturally occurring curvature of film at scanning, giving rise to a maximum height 1 to 2 mm above the scan plane, may introduce dose errors of 1% to 4%, and considerably reduce gamma evaluation passing rates when comparing film‐measured doses with treatment planning system‐calculated dose distributions, a common application of film dosimetry in radiotherapy. The use of a triple‐channel dosimetry algorithm appeared to mitigate the error due to film curvature compared to conventional single‐channel film dosimetry. The change in pixel value and calibrated reported dose with film curling or height above the scanner plate may be due to variations in illumination characteristics, optical disturbances, or a Callier‐type effect. There is a clear requirement for physically flat films at scanning to avoid the introduction of a substantial error source in film dosimetry. Particularly for small film samples, a compression glass plate above the film is recommended to ensure flat‐film scanning. This effect has been overlooked to date in the literature. PACS numbers: 87.55.Qr, 87.56.bg, 87.55.km
format Online
Article
Text
id pubmed-5690100
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56901002018-04-02 Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning Palmer, Antony L. Bradley, David A. Nisbet, Andrew J Appl Clin Med Phys Radiation Measurements This work considers a previously overlooked uncertainty present in film dosimetry which results from moderate curvature of films during the scanning process. Small film samples are particularly susceptible to film curling which may be undetected or deemed insignificant. In this study, we consider test cases with controlled induced curvature of film and with film raised horizontally above the scanner plate. We also evaluate the difference in scans of a film irradiated with a typical brachytherapy dose distribution with the film naturally curved and with the film held flat on the scanner. Typical naturally occurring curvature of film at scanning, giving rise to a maximum height 1 to 2 mm above the scan plane, may introduce dose errors of 1% to 4%, and considerably reduce gamma evaluation passing rates when comparing film‐measured doses with treatment planning system‐calculated dose distributions, a common application of film dosimetry in radiotherapy. The use of a triple‐channel dosimetry algorithm appeared to mitigate the error due to film curvature compared to conventional single‐channel film dosimetry. The change in pixel value and calibrated reported dose with film curling or height above the scanner plate may be due to variations in illumination characteristics, optical disturbances, or a Callier‐type effect. There is a clear requirement for physically flat films at scanning to avoid the introduction of a substantial error source in film dosimetry. Particularly for small film samples, a compression glass plate above the film is recommended to ensure flat‐film scanning. This effect has been overlooked to date in the literature. PACS numbers: 87.55.Qr, 87.56.bg, 87.55.km John Wiley and Sons Inc. 2015-03-08 /pmc/articles/PMC5690100/ /pubmed/26103181 http://dx.doi.org/10.1120/jacmp.v16i2.5141 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Measurements
Palmer, Antony L.
Bradley, David A.
Nisbet, Andrew
Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
title Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
title_full Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
title_fullStr Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
title_full_unstemmed Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
title_short Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
title_sort evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
topic Radiation Measurements
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690100/
https://www.ncbi.nlm.nih.gov/pubmed/26103181
http://dx.doi.org/10.1120/jacmp.v16i2.5141
work_keys_str_mv AT palmerantonyl evaluationandmitigationofpotentialerrorsinradiochromicfilmdosimetryduetofilmcurvatureatscanning
AT bradleydavida evaluationandmitigationofpotentialerrorsinradiochromicfilmdosimetryduetofilmcurvatureatscanning
AT nisbetandrew evaluationandmitigationofpotentialerrorsinradiochromicfilmdosimetryduetofilmcurvatureatscanning