Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement

Radiographic film dosimetry suffers from its energy dependence in proton dosimetry. This study sought to develop a method of measuring proton beams by the film and to evaluate film response to proton beams for the constancy check of depth dose (DD). It also evaluated the film for profile measurement...

Descripción completa

Detalles Bibliográficos
Autores principales: Yeo, Inhwan J., Teran, Anthony, Ghebremedhin, Abiel, Johnson, Matt, Patyal, Baldev
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690120/
https://www.ncbi.nlm.nih.gov/pubmed/26103499
http://dx.doi.org/10.1120/jacmp.v16i3.5402
_version_ 1783279534380941312
author Yeo, Inhwan J.
Teran, Anthony
Ghebremedhin, Abiel
Johnson, Matt
Patyal, Baldev
author_facet Yeo, Inhwan J.
Teran, Anthony
Ghebremedhin, Abiel
Johnson, Matt
Patyal, Baldev
author_sort Yeo, Inhwan J.
collection PubMed
description Radiographic film dosimetry suffers from its energy dependence in proton dosimetry. This study sought to develop a method of measuring proton beams by the film and to evaluate film response to proton beams for the constancy check of depth dose (DD). It also evaluated the film for profile measurements. To achieve this goal, from DDs measured by film and ion chamber (IC), calibration factors (ratios of dose measured by IC to film responses) as a function of depth in a phantom were obtained. These factors imply variable slopes (with proton energy and depth) of linear characteristic curves that relate film response to dose. We derived a calibration method that enables utilization of the factors for acquisition of dose from film density measured at later dates by adapting to a potentially altered processor condition. To test this model, the characteristic curve was obtained by using EDR2 film and in‐phantom film dosimetry in parallel with a 149.65 MeV proton beam, using the method. An additional validation of the model was performed by concurrent film and IC measurement perpendicular to the beam at various depths. Beam profile measurements by the film were also evaluated at the center of beam modulation. In order to interpret and ascertain the film dosimetry, Monte Carlos simulation of the beam was performed, calculating the proton fluence spectrum along depths and off‐axis distances. By multiplying respective stopping powers to the spectrum, doses to film and water were calculated. The ratio of film dose to water dose was evaluated. Results are as follows. The characteristic curve proved the assumed linearity. The measured DD approached that of IC, but near the end of the spread‐out Bragg peak (SOBP), a spurious peak was observed due to the mismatch of distal edge between the calibration and measurement films. The width of SOBP and the proximal edge were both reproducible within a maximum of 5 mm; the distal edge was reproducible within 1 mm. At 5 cm depth, the dose was reproducible within 10%. These large discrepancies were identified to have been contributed by film processor uncertainty across a layer of film and the misalignment of film edge to the frontal phantom surface. The deviations could drop from 5 to 2 mm in SOBP and from 10% to 4.5% at 5 cm depth in a well‐controlled processor condition (i.e., warm up). In addition to the validation of the calibration method done by the DD measurements, the concurrent film and IC measurement independently validated the model by showing the constancy of depth‐dependent calibration factors. For profile measurement, the film showed good agreement with ion chamber measurement. In agreement with the experimental findings, computationally obtained ratio of film dose to water dose assisted understanding of the trend of the film response by revealing relatively large and small variances of the response for DD and beam profile measurements, respectively. Conclusions are as follows. For proton beams, radiographic film proved to offer accurate beam profile measurements. The adaptive calibration method proposed in this study was validated. Using the method, film dosimetry could offer reasonably accurate DD constancy checks, when provided with a well‐controlled processor condition. Although the processor warming up can promote a uniform processing across a single layer of the film, the processing remains as a challenge. PACS number: 87
format Online
Article
Text
id pubmed-5690120
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56901202018-04-02 Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement Yeo, Inhwan J. Teran, Anthony Ghebremedhin, Abiel Johnson, Matt Patyal, Baldev J Appl Clin Med Phys Radiation Measurements Radiographic film dosimetry suffers from its energy dependence in proton dosimetry. This study sought to develop a method of measuring proton beams by the film and to evaluate film response to proton beams for the constancy check of depth dose (DD). It also evaluated the film for profile measurements. To achieve this goal, from DDs measured by film and ion chamber (IC), calibration factors (ratios of dose measured by IC to film responses) as a function of depth in a phantom were obtained. These factors imply variable slopes (with proton energy and depth) of linear characteristic curves that relate film response to dose. We derived a calibration method that enables utilization of the factors for acquisition of dose from film density measured at later dates by adapting to a potentially altered processor condition. To test this model, the characteristic curve was obtained by using EDR2 film and in‐phantom film dosimetry in parallel with a 149.65 MeV proton beam, using the method. An additional validation of the model was performed by concurrent film and IC measurement perpendicular to the beam at various depths. Beam profile measurements by the film were also evaluated at the center of beam modulation. In order to interpret and ascertain the film dosimetry, Monte Carlos simulation of the beam was performed, calculating the proton fluence spectrum along depths and off‐axis distances. By multiplying respective stopping powers to the spectrum, doses to film and water were calculated. The ratio of film dose to water dose was evaluated. Results are as follows. The characteristic curve proved the assumed linearity. The measured DD approached that of IC, but near the end of the spread‐out Bragg peak (SOBP), a spurious peak was observed due to the mismatch of distal edge between the calibration and measurement films. The width of SOBP and the proximal edge were both reproducible within a maximum of 5 mm; the distal edge was reproducible within 1 mm. At 5 cm depth, the dose was reproducible within 10%. These large discrepancies were identified to have been contributed by film processor uncertainty across a layer of film and the misalignment of film edge to the frontal phantom surface. The deviations could drop from 5 to 2 mm in SOBP and from 10% to 4.5% at 5 cm depth in a well‐controlled processor condition (i.e., warm up). In addition to the validation of the calibration method done by the DD measurements, the concurrent film and IC measurement independently validated the model by showing the constancy of depth‐dependent calibration factors. For profile measurement, the film showed good agreement with ion chamber measurement. In agreement with the experimental findings, computationally obtained ratio of film dose to water dose assisted understanding of the trend of the film response by revealing relatively large and small variances of the response for DD and beam profile measurements, respectively. Conclusions are as follows. For proton beams, radiographic film proved to offer accurate beam profile measurements. The adaptive calibration method proposed in this study was validated. Using the method, film dosimetry could offer reasonably accurate DD constancy checks, when provided with a well‐controlled processor condition. Although the processor warming up can promote a uniform processing across a single layer of the film, the processing remains as a challenge. PACS number: 87 John Wiley and Sons Inc. 2015-05-08 /pmc/articles/PMC5690120/ /pubmed/26103499 http://dx.doi.org/10.1120/jacmp.v16i3.5402 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Measurements
Yeo, Inhwan J.
Teran, Anthony
Ghebremedhin, Abiel
Johnson, Matt
Patyal, Baldev
Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
title Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
title_full Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
title_fullStr Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
title_full_unstemmed Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
title_short Radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
title_sort radiographic film dosimetry of proton beams for depth‐dose constancy check and beam profile measurement
topic Radiation Measurements
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690120/
https://www.ncbi.nlm.nih.gov/pubmed/26103499
http://dx.doi.org/10.1120/jacmp.v16i3.5402
work_keys_str_mv AT yeoinhwanj radiographicfilmdosimetryofprotonbeamsfordepthdoseconstancycheckandbeamprofilemeasurement
AT terananthony radiographicfilmdosimetryofprotonbeamsfordepthdoseconstancycheckandbeamprofilemeasurement
AT ghebremedhinabiel radiographicfilmdosimetryofprotonbeamsfordepthdoseconstancycheckandbeamprofilemeasurement
AT johnsonmatt radiographicfilmdosimetryofprotonbeamsfordepthdoseconstancycheckandbeamprofilemeasurement
AT patyalbaldev radiographicfilmdosimetryofprotonbeamsfordepthdoseconstancycheckandbeamprofilemeasurement