Evaluation of monitor unit calculation based on measurement and calculation with a simplified Monte Carlo method for passive beam delivery system in proton beam therapy

Calibrating the dose per monitor unit (DMU) for individual patients is important to deliver the prescribed dose in radiation therapy. We have developed a DMU calculation method combining measurement data and calculation with a simplified Monte Carlo method for the double scattering system in proton...

Descripción completa

Detalles Bibliográficos
Autores principales: Hotta, Kenji, Kohno, Ryosuke, Nagafuchi, Kohsuke, Yamaguchi, Hidenori, Tansho, Ryohei, Takada, Yoshihisa, Akimoto, Tetsuo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690152/
https://www.ncbi.nlm.nih.gov/pubmed/26699303
http://dx.doi.org/10.1120/jacmp.v16i5.5419
Descripción
Sumario:Calibrating the dose per monitor unit (DMU) for individual patients is important to deliver the prescribed dose in radiation therapy. We have developed a DMU calculation method combining measurement data and calculation with a simplified Monte Carlo method for the double scattering system in proton beam therapy at the National Cancer Center Hospital East in Japan. The DMU calculation method determines the clinical DMU by the multiplication of three factors: a beam spreading device factor [Formula: see text] , a patient‐specific device factor [Formula: see text] , and a field‐size correction factor [Formula: see text]. We compared the calculated and the measured DMU for 75 dose fields in clinical cases. The calculated DMUs were in agreement with measurements in [Formula: see text] for all of 25 fields in prostate cancer cases, and in [Formula: see text] for 94% of 50 fields in head and neck (H&N) and lung cancer cases, including irregular shape fields and small fields. Although the [Formula: see text] in the DMU calculations is dominant as expected, we found that the patient‐specific device factor and field‐size correction also contribute significantly to the calculated DMU. This DMU calculation method will be able to substitute the conventional DMU measurement for the majority of clinical cases with a reasonable calculation time required for clinical use. PACS number: 87.55.kh