Cargando…

Comparison of pediatric radiation dose and vessel visibility on angiographic systems using piglets as a surrogate: antiscatter grid removal vs. lower detector air kerma settings with a grid — a preclinical investigation

The purpose of this study was to reduce pediatric doses while maintaining or improving image quality scores without removing the grid from X‐ray beam. This study was approved by the Institutional Animal Care and Use Committee. Three piglets (5, 14, and 20 kg) were imaged using six different selectab...

Descripción completa

Detalles Bibliográficos
Autores principales: Strauss, Keith J., Racadio, John M., Abruzzo, Todd A., Johnson, Neil D., Patel, Manish N., Kukreja, Kamlesh U., den Hartog, Mark. J. H., Hoornaert, Bart P.A., Nachabe, Rami A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690159/
https://www.ncbi.nlm.nih.gov/pubmed/26699297
http://dx.doi.org/10.1120/jacmp.v16i5.5379
Descripción
Sumario:The purpose of this study was to reduce pediatric doses while maintaining or improving image quality scores without removing the grid from X‐ray beam. This study was approved by the Institutional Animal Care and Use Committee. Three piglets (5, 14, and 20 kg) were imaged using six different selectable detector air kerma ([Formula: see text]) per frame values (100%, 70%, 50%, 35%, 25%, 17.5%) with and without the grid. Number of distal branches visualized with diagnostic confidence relative to the injected vessel defined image quality score. Five pediatric interventional radiologists evaluated all images. Image quality score and piglet [Formula: see text] were statistically compared using analysis of variance and receiver operating curve analysis to define the preferred dose setting and use of grid for a visibility of 2nd and 3rd order vessel branches. Grid removal reduced both dose to subject and imaging quality by 26%. Third order branches could only be visualized with the grid present; 100% detector [Formula: see text] was required for smallest pig, while 70% detector [Formula: see text] was adequate for the two larger pigs. Second order branches could be visualized with grid at 17.5% detector [Formula: see text] for all three pig sizes. Without the grid, 50%, 35%, and 35% detector [Formula: see text] were required for smallest to largest pig, respectively. Grid removal reduces both dose and image quality score. Image quality scores can be maintained with less dose to subject with the grid in the beam as opposed to removed. Smaller anatomy requires more dose to the detector to achieve the same image quality score. PACS numbers: 87.53.Bn, 87.57.N‐, 87.57.cj, 87.59.cf, 87.59.Dj