Cargando…
On the sensitivity of TG‐119 and IROC credentialing to TPS commissioning errors
We investigate the sensitivity of IMRT commissioning using the TG‐119 C‐shape phantom and credentialing with the IROC head and neck phantom to treatment planning system commissioning errors. We introduced errors into the various aspects of the commissioning process for a 6X photon energy modeled usi...
Autores principales: | McVicker, Drew, Yin, Fang‐Fang, Adamson, Justus D. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690193/ https://www.ncbi.nlm.nih.gov/pubmed/26894330 http://dx.doi.org/10.1120/jacmp.v17i1.5452 |
Ejemplares similares
-
Impact of the MLC leaf‐tip model in a commercial TPS: Dose calculation limitations and IROC‐H phantom failures
por: Koger, Brandon, et al.
Publicado: (2020) -
IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG‐119 protocol cases
por: Wen, Ning, et al.
Publicado: (2014) -
Application of AAPM TG 119 to volumetric arc therapy (VMAT)
por: Mynampati, Dinesh Kumar, et al.
Publicado: (2012) -
Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience
por: Chang, Zheng, et al.
Publicado: (2012) -
Evaluation of six TPS algorithms in computing entrance and exit doses
por: Tan, Yun I., et al.
Publicado: (2014)