Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria

The original helical ArcCHECK (AC) diode array and associated software for 3D measurement‐guided dose reconstruction were characterized and validated; however, recent design changes to the AC required that the subject be revisited. The most important AC change starting in 2014 was a significant redu...

Descripción completa

Detalles Bibliográficos
Autores principales: Ahmed, Saeed, Nelms, Benjamin, Kozelka, Jakub, Zhang, Geoffrey, Moros, Eduardo, Feygelman, Vladimir
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690493/
https://www.ncbi.nlm.nih.gov/pubmed/27929491
http://dx.doi.org/10.1120/jacmp.v17i6.6414
_version_ 1783279619067084800
author Ahmed, Saeed
Nelms, Benjamin
Kozelka, Jakub
Zhang, Geoffrey
Moros, Eduardo
Feygelman, Vladimir
author_facet Ahmed, Saeed
Nelms, Benjamin
Kozelka, Jakub
Zhang, Geoffrey
Moros, Eduardo
Feygelman, Vladimir
author_sort Ahmed, Saeed
collection PubMed
description The original helical ArcCHECK (AC) diode array and associated software for 3D measurement‐guided dose reconstruction were characterized and validated; however, recent design changes to the AC required that the subject be revisited. The most important AC change starting in 2014 was a significant reduction in the overresponse of diodes to scattered radiation outside of the direct beam, accomplished by reducing the amount of high‐Z materials adjacent to the diodes. This change improved the diode measurement accuracy, but in the process invalidated the dose reconstruction models that were assembled based on measured data acquired with the older version of the AC. A correction mechanism was introduced in the reconstruction software (3DVH) to accommodate this and potential future design changes without requiring updating model parameters. For each permutation of AC serial number and beam model, the user can define in 3DVH a single correction factor which will be used to compensate for the difference in the out‐of‐field response between the new and original AC designs. The exact value can be determined by minimizing the dose‐difference with an ionization chamber or another independent dosimeter. A single value of 1.17, corresponding to the maximum measured out‐of‐field response difference between the new and old AC, provided satisfactory results for all studied energies (6X, 15X, and flattening filter‐free 10XFFF). A library of standard cases recommended by the AAPM TG‐244 Report was used for reconstructed dose verification. The overall difference between reconstructed dose and an ion chamber in a water‐equivalent phantom in the targets was [Formula: see text] (1 SD). The reconstructed dose on a homogeneous phantom was also compared to a biplanar diode dosimeter ([Formula: see text]) using gamma analysis with [Formula: see text] cutoff criteria. The mean agreement rate was [Formula: see text]. For the plans common with the previous comparison, the mean agreement rate was [Formula: see text] , essentially unchanged. We conclude that the proposed software modification adequately addresses the change in the dosimeter response. PACS number(s): 87.55Qr
format Online
Article
Text
id pubmed-5690493
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56904932018-04-02 Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria Ahmed, Saeed Nelms, Benjamin Kozelka, Jakub Zhang, Geoffrey Moros, Eduardo Feygelman, Vladimir J Appl Clin Med Phys Radiation Oncology Physics The original helical ArcCHECK (AC) diode array and associated software for 3D measurement‐guided dose reconstruction were characterized and validated; however, recent design changes to the AC required that the subject be revisited. The most important AC change starting in 2014 was a significant reduction in the overresponse of diodes to scattered radiation outside of the direct beam, accomplished by reducing the amount of high‐Z materials adjacent to the diodes. This change improved the diode measurement accuracy, but in the process invalidated the dose reconstruction models that were assembled based on measured data acquired with the older version of the AC. A correction mechanism was introduced in the reconstruction software (3DVH) to accommodate this and potential future design changes without requiring updating model parameters. For each permutation of AC serial number and beam model, the user can define in 3DVH a single correction factor which will be used to compensate for the difference in the out‐of‐field response between the new and original AC designs. The exact value can be determined by minimizing the dose‐difference with an ionization chamber or another independent dosimeter. A single value of 1.17, corresponding to the maximum measured out‐of‐field response difference between the new and old AC, provided satisfactory results for all studied energies (6X, 15X, and flattening filter‐free 10XFFF). A library of standard cases recommended by the AAPM TG‐244 Report was used for reconstructed dose verification. The overall difference between reconstructed dose and an ion chamber in a water‐equivalent phantom in the targets was [Formula: see text] (1 SD). The reconstructed dose on a homogeneous phantom was also compared to a biplanar diode dosimeter ([Formula: see text]) using gamma analysis with [Formula: see text] cutoff criteria. The mean agreement rate was [Formula: see text]. For the plans common with the previous comparison, the mean agreement rate was [Formula: see text] , essentially unchanged. We conclude that the proposed software modification adequately addresses the change in the dosimeter response. PACS number(s): 87.55Qr John Wiley and Sons Inc. 2016-11-08 /pmc/articles/PMC5690493/ /pubmed/27929491 http://dx.doi.org/10.1120/jacmp.v17i6.6414 Text en © 2016 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Ahmed, Saeed
Nelms, Benjamin
Kozelka, Jakub
Zhang, Geoffrey
Moros, Eduardo
Feygelman, Vladimir
Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria
title Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria
title_full Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria
title_fullStr Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria
title_full_unstemmed Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria
title_short Validation of an improved helical diode array and dose reconstruction software using TG‐244 datasets and stringent dose comparison criteria
title_sort validation of an improved helical diode array and dose reconstruction software using tg‐244 datasets and stringent dose comparison criteria
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690493/
https://www.ncbi.nlm.nih.gov/pubmed/27929491
http://dx.doi.org/10.1120/jacmp.v17i6.6414
work_keys_str_mv AT ahmedsaeed validationofanimprovedhelicaldiodearrayanddosereconstructionsoftwareusingtg244datasetsandstringentdosecomparisoncriteria
AT nelmsbenjamin validationofanimprovedhelicaldiodearrayanddosereconstructionsoftwareusingtg244datasetsandstringentdosecomparisoncriteria
AT kozelkajakub validationofanimprovedhelicaldiodearrayanddosereconstructionsoftwareusingtg244datasetsandstringentdosecomparisoncriteria
AT zhanggeoffrey validationofanimprovedhelicaldiodearrayanddosereconstructionsoftwareusingtg244datasetsandstringentdosecomparisoncriteria
AT moroseduardo validationofanimprovedhelicaldiodearrayanddosereconstructionsoftwareusingtg244datasetsandstringentdosecomparisoncriteria
AT feygelmanvladimir validationofanimprovedhelicaldiodearrayanddosereconstructionsoftwareusingtg244datasetsandstringentdosecomparisoncriteria