Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction

To evaluate the feasibility of radiation dose reduction using model‐based iterative reconstruction (MBIR) for evaluating the ureters and vasculature in a phantom, a tissue‐equivalent CT dose phantom was scanned using a 64‐channel CT scanner. Tubes of varying diameters filled with different dilutions...

Descripción completa

Detalles Bibliográficos
Autores principales: Page, Leland, Wei, Wei, Kundra, Vikas, Rong, X. John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690497/
https://www.ncbi.nlm.nih.gov/pubmed/27929506
http://dx.doi.org/10.1120/jacmp.v17i6.6184
_version_ 1783279620033871872
author Page, Leland
Wei, Wei
Kundra, Vikas
Rong, X. John
author_facet Page, Leland
Wei, Wei
Kundra, Vikas
Rong, X. John
author_sort Page, Leland
collection PubMed
description To evaluate the feasibility of radiation dose reduction using model‐based iterative reconstruction (MBIR) for evaluating the ureters and vasculature in a phantom, a tissue‐equivalent CT dose phantom was scanned using a 64‐channel CT scanner. Tubes of varying diameters filled with different dilutions of a contrast agent, simulating ureters or vessels, were inserted into the center of the phantom. Each combination was scanned using an existing renal protocol at 140 kVp or 120 kVp, yielding a display volumetric CT dose index ([Formula: see text]) of 24 mGy. The scans were repeated using reduced scan techniques to achieve lower radiation doses down to 0.8 mGy. The images were reconstructed using filtered back‐projection (FBP) and model‐based iterative reconstruction (MBIR). The noise and contrast‐to‐noise ratio (CNR) was measured for each contrast object. Comparisons between the two reconstruction methods at different dose levels were evaluated using a factorial design. At each [Formula: see text] the measured image noise was lower using MBIR compared to FBP ([Formula: see text]). At low doses, the percent change in measured image noise between FBP and MBIR was larger. For the 12 mm object simulating a ureter or large vessel with an HU of 600, the measured CNR using MBIR at a [Formula: see text] of 1.7 mGy was greater than the CNR of FBP at a [Formula: see text] of 24 mGy ([Formula: see text]). For the 5 mm object simulating a medium‐sized vessel with a HU of 250, the measured CNR using MBIR at a [Formula: see text] of 1.7 mGy was equivalent to that of FBP at a [Formula: see text] of 24 mGy. For the 2 mm, 100 HU object simulating a small vessel, the measured CNR using MBIR at a [Formula: see text] of 1.7 mGy was equivalent to that of FBP at a [Formula: see text] of 24 mGy. Low‐dose (3.6 mGy) CT imaging of vasculature and ureter phantoms using MBIR results in similar noise and CNR compared to FBP at approximately one‐sixth the dose. This suggests that, using MBIR, a one milliSievert exam of the ureters and vasculature may be clinically possible whilst still maintaining adequate image quality PACS number(s): 87.57.‐s, 87.57.Q‐, 87.57.C‐, 87.57.nf, 87.57.cj, 87.57.cm
format Online
Article
Text
id pubmed-5690497
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56904972018-04-02 Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction Page, Leland Wei, Wei Kundra, Vikas Rong, X. John J Appl Clin Med Phys Medical Imaging To evaluate the feasibility of radiation dose reduction using model‐based iterative reconstruction (MBIR) for evaluating the ureters and vasculature in a phantom, a tissue‐equivalent CT dose phantom was scanned using a 64‐channel CT scanner. Tubes of varying diameters filled with different dilutions of a contrast agent, simulating ureters or vessels, were inserted into the center of the phantom. Each combination was scanned using an existing renal protocol at 140 kVp or 120 kVp, yielding a display volumetric CT dose index ([Formula: see text]) of 24 mGy. The scans were repeated using reduced scan techniques to achieve lower radiation doses down to 0.8 mGy. The images were reconstructed using filtered back‐projection (FBP) and model‐based iterative reconstruction (MBIR). The noise and contrast‐to‐noise ratio (CNR) was measured for each contrast object. Comparisons between the two reconstruction methods at different dose levels were evaluated using a factorial design. At each [Formula: see text] the measured image noise was lower using MBIR compared to FBP ([Formula: see text]). At low doses, the percent change in measured image noise between FBP and MBIR was larger. For the 12 mm object simulating a ureter or large vessel with an HU of 600, the measured CNR using MBIR at a [Formula: see text] of 1.7 mGy was greater than the CNR of FBP at a [Formula: see text] of 24 mGy ([Formula: see text]). For the 5 mm object simulating a medium‐sized vessel with a HU of 250, the measured CNR using MBIR at a [Formula: see text] of 1.7 mGy was equivalent to that of FBP at a [Formula: see text] of 24 mGy. For the 2 mm, 100 HU object simulating a small vessel, the measured CNR using MBIR at a [Formula: see text] of 1.7 mGy was equivalent to that of FBP at a [Formula: see text] of 24 mGy. Low‐dose (3.6 mGy) CT imaging of vasculature and ureter phantoms using MBIR results in similar noise and CNR compared to FBP at approximately one‐sixth the dose. This suggests that, using MBIR, a one milliSievert exam of the ureters and vasculature may be clinically possible whilst still maintaining adequate image quality PACS number(s): 87.57.‐s, 87.57.Q‐, 87.57.C‐, 87.57.nf, 87.57.cj, 87.57.cm John Wiley and Sons Inc. 2016-11-08 /pmc/articles/PMC5690497/ /pubmed/27929506 http://dx.doi.org/10.1120/jacmp.v17i6.6184 Text en © 2016 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Page, Leland
Wei, Wei
Kundra, Vikas
Rong, X. John
Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction
title Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction
title_full Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction
title_fullStr Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction
title_full_unstemmed Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction
title_short Dose reduction in CT urography and vasculature phantom studies using model‐based iterative reconstruction
title_sort dose reduction in ct urography and vasculature phantom studies using model‐based iterative reconstruction
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690497/
https://www.ncbi.nlm.nih.gov/pubmed/27929506
http://dx.doi.org/10.1120/jacmp.v17i6.6184
work_keys_str_mv AT pageleland dosereductionincturographyandvasculaturephantomstudiesusingmodelbasediterativereconstruction
AT weiwei dosereductionincturographyandvasculaturephantomstudiesusingmodelbasediterativereconstruction
AT kundravikas dosereductionincturographyandvasculaturephantomstudiesusingmodelbasediterativereconstruction
AT rongxjohn dosereductionincturographyandvasculaturephantomstudiesusingmodelbasediterativereconstruction