Improvement in CT image resolution due to the use of focal spot deflection and increased sampling

When patient anatomy is positioned away from a CT scanner's isocenter, scans of limited diagnostic value may result. Yet in some cases, positioning of patient anatomy far from isocenter is unavoidable. This study examines the effect of position and reconstruction algorithm on image resolution a...

Descripción completa

Detalles Bibliográficos
Autores principales: Rubert, Nicholas, Szczykutowicz, Timothy, Ranallo, Frank
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690917/
https://www.ncbi.nlm.nih.gov/pubmed/27167276
http://dx.doi.org/10.1120/jacmp.v17i3.6039
_version_ 1783279693240205312
author Rubert, Nicholas
Szczykutowicz, Timothy
Ranallo, Frank
author_facet Rubert, Nicholas
Szczykutowicz, Timothy
Ranallo, Frank
author_sort Rubert, Nicholas
collection PubMed
description When patient anatomy is positioned away from a CT scanner's isocenter, scans of limited diagnostic value may result. Yet in some cases, positioning of patient anatomy far from isocenter is unavoidable. This study examines the effect of position and reconstruction algorithm on image resolution achieved by a CT scanner operating in a high resolution (HR) scan mode which incorporates focal spot deflection and acquires an increased number of projections per rotation. Images of a metal bead contained in a phantom were acquired on a GE CT750 HD scanner with multiple reconstruction algorithms, in the normal and HR scan mode, and at two positions, scanner isocenter and 15 cm directly above isocenter. The images of the metal bead yielded two‐dimensional point spread functions which were averaged along two perpendicular directions to yield line spread functions. Fourier transforms of the line spread functions yielded radial and azimuthal modulation transfer functions (MTFs). At isocenter, the radial and azimuthal MTFs were averaged. MTF improvement depended on image position and modulation direction. The results from a single algorithm, Edge, can be generalized to other algorithms. At isocenter, the 10% MTF cutoff was 14.4 cycles/cm in normal and HR mode. At 15 cm above isocenter, the 10% cutoff was 6.0 and 8.5 cycles/cm for the azimuthal and radial MTFs in normal mode. In HR mode, the azimuthal and radial MTF 10% cutoff was 8.3 and 10.3 cycles/cm. Our results indicate that the best image resolution is achieved at scanner isocenter and that the azimuthal resolution degrades more significantly than the radial resolution. For the GE CT750 HD CT scanner, the resolution is significantly enhanced by the HR scan mode away from scanner isocenter, and the use of the HR scan mode has much more of an impact on image resolution away from isocenter than the choice of algorithm. PACS number(s): 87.57.Q‐
format Online
Article
Text
id pubmed-5690917
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56909172018-04-02 Improvement in CT image resolution due to the use of focal spot deflection and increased sampling Rubert, Nicholas Szczykutowicz, Timothy Ranallo, Frank J Appl Clin Med Phys Medical Imaging When patient anatomy is positioned away from a CT scanner's isocenter, scans of limited diagnostic value may result. Yet in some cases, positioning of patient anatomy far from isocenter is unavoidable. This study examines the effect of position and reconstruction algorithm on image resolution achieved by a CT scanner operating in a high resolution (HR) scan mode which incorporates focal spot deflection and acquires an increased number of projections per rotation. Images of a metal bead contained in a phantom were acquired on a GE CT750 HD scanner with multiple reconstruction algorithms, in the normal and HR scan mode, and at two positions, scanner isocenter and 15 cm directly above isocenter. The images of the metal bead yielded two‐dimensional point spread functions which were averaged along two perpendicular directions to yield line spread functions. Fourier transforms of the line spread functions yielded radial and azimuthal modulation transfer functions (MTFs). At isocenter, the radial and azimuthal MTFs were averaged. MTF improvement depended on image position and modulation direction. The results from a single algorithm, Edge, can be generalized to other algorithms. At isocenter, the 10% MTF cutoff was 14.4 cycles/cm in normal and HR mode. At 15 cm above isocenter, the 10% cutoff was 6.0 and 8.5 cycles/cm for the azimuthal and radial MTFs in normal mode. In HR mode, the azimuthal and radial MTF 10% cutoff was 8.3 and 10.3 cycles/cm. Our results indicate that the best image resolution is achieved at scanner isocenter and that the azimuthal resolution degrades more significantly than the radial resolution. For the GE CT750 HD CT scanner, the resolution is significantly enhanced by the HR scan mode away from scanner isocenter, and the use of the HR scan mode has much more of an impact on image resolution away from isocenter than the choice of algorithm. PACS number(s): 87.57.Q‐ John Wiley and Sons Inc. 2016-05-08 /pmc/articles/PMC5690917/ /pubmed/27167276 http://dx.doi.org/10.1120/jacmp.v17i3.6039 Text en © 2016 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Rubert, Nicholas
Szczykutowicz, Timothy
Ranallo, Frank
Improvement in CT image resolution due to the use of focal spot deflection and increased sampling
title Improvement in CT image resolution due to the use of focal spot deflection and increased sampling
title_full Improvement in CT image resolution due to the use of focal spot deflection and increased sampling
title_fullStr Improvement in CT image resolution due to the use of focal spot deflection and increased sampling
title_full_unstemmed Improvement in CT image resolution due to the use of focal spot deflection and increased sampling
title_short Improvement in CT image resolution due to the use of focal spot deflection and increased sampling
title_sort improvement in ct image resolution due to the use of focal spot deflection and increased sampling
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690917/
https://www.ncbi.nlm.nih.gov/pubmed/27167276
http://dx.doi.org/10.1120/jacmp.v17i3.6039
work_keys_str_mv AT rubertnicholas improvementinctimageresolutionduetotheuseoffocalspotdeflectionandincreasedsampling
AT szczykutowicztimothy improvementinctimageresolutionduetotheuseoffocalspotdeflectionandincreasedsampling
AT ranallofrank improvementinctimageresolutionduetotheuseoffocalspotdeflectionandincreasedsampling