Cargando…

Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators

Flattening filter‐free (FFF) beams are becoming the preferred beam type for stereotactic radiosurgery (SRS) and stereotactic ablative radiation therapy (SABR), as they enable an increase in dose rate and a decrease in treatment time. This work assesses the effects of the flattening filter on small f...

Descripción completa

Detalles Bibliográficos
Autores principales: Tyler, Madelaine K., Liu, Paul Z.Y., Lee, Christopher, McKenzie, David R., Suchowerska, Natalka
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690940/
https://www.ncbi.nlm.nih.gov/pubmed/27167280
http://dx.doi.org/10.1120/jacmp.v17i3.6059
Descripción
Sumario:Flattening filter‐free (FFF) beams are becoming the preferred beam type for stereotactic radiosurgery (SRS) and stereotactic ablative radiation therapy (SABR), as they enable an increase in dose rate and a decrease in treatment time. This work assesses the effects of the flattening filter on small field output factors for 6 MV beams generated by both Elekta and Varian linear accelerators, and determines differences between detector response in flattened (FF) and FFF beams. Relative output factors were measured with a range of detectors (diodes, ionization chambers, radiochromic film, and microDiamond) and referenced to the relative output factors measured with an air core fiber optic dosimeter (FOD), a scintillation dosimeter developed at Chris O'Brien Lifehouse, Sydney. Small field correction factors were generated for both FF and FFF beams. Diode measured detector response was compared with a recently published mathematical relation to predict diode response corrections in small fields. The effect of flattening filter removal on detector response was quantified using a ratio of relative detector responses in FFF and FF fields for the same field size. The removal of the flattening filter was found to have a small but measurable effect on ionization chamber response with maximum deviations of less than [Formula: see text] across all field sizes measured. Solid‐state detectors showed an increased dependence on the flattening filter of up to [Formula: see text]. Measured diode response was within [Formula: see text] of the published mathematical relation for all fields up to 30 mm, independent of linac type and presence or absence of a flattening filter. For 6 MV beams, detector correction factors between FFF and FF beams are interchangeable for a linac between FF and FFF modes, providing that an additional uncertainty of up to [Formula: see text] is accepted. PACS number(s): 87.55.km, 87.56.bd, 87.56.Da