Cargando…

CT radiation profile width measurement using CR imaging plate raw data

This technical note demonstrates computed tomography (CT) radiation profile measurement using computed radiography (CR) imaging plate raw data showing it is possible to perform the CT collimation width measurement using a single scan without saturating the imaging plate. Previously described methods...

Descripción completa

Detalles Bibliográficos
Autores principales: Bjarnason, Thorarin A, Yang, Chang‐Ying Joseph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690993/
https://www.ncbi.nlm.nih.gov/pubmed/26699559
http://dx.doi.org/10.1120/jacmp.v16i6.5512
_version_ 1783279701679144960
author Bjarnason, Thorarin A
Yang, Chang‐Ying Joseph
author_facet Bjarnason, Thorarin A
Yang, Chang‐Ying Joseph
author_sort Bjarnason, Thorarin A
collection PubMed
description This technical note demonstrates computed tomography (CT) radiation profile measurement using computed radiography (CR) imaging plate raw data showing it is possible to perform the CT collimation width measurement using a single scan without saturating the imaging plate. Previously described methods require careful adjustments to the CR reader settings in order to avoid signal clipping in the CR processed image. CT radiation profile measurements were taken as part of routine quality control on 14 CT scanners from four vendors. CR cassettes were placed on the CT scanner bed, raised to isocenter, and leveled. Axial scans were taken at all available collimations, advancing the cassette for each scan. The CR plates were processed and raw CR data were analyzed using MATLAB scripts to measure collimation widths. The raw data approach was compared with previously established methodology. The quality control analysis scripts are released as open source using creative commons licensing. A log‐linear relationship was found between raw pixel value and air kerma, and raw data collimation width measurements were in agreement with CR‐processed, bit‐reduced data, using previously described methodology. The raw data approach, with intrinsically wider dynamic range, allows improved measurement flexibility and precision. As a result, we demonstrate a methodology for CT collimation width measurements using a single CT scan and without the need for CR scanning parameter adjustments which is more convenient for routine quality control work. PACS numbers: 87.57.Q‐, 87.59.bd, 87.57.uq
format Online
Article
Text
id pubmed-5690993
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56909932018-04-02 CT radiation profile width measurement using CR imaging plate raw data Bjarnason, Thorarin A Yang, Chang‐Ying Joseph J Appl Clin Med Phys Technical Notes This technical note demonstrates computed tomography (CT) radiation profile measurement using computed radiography (CR) imaging plate raw data showing it is possible to perform the CT collimation width measurement using a single scan without saturating the imaging plate. Previously described methods require careful adjustments to the CR reader settings in order to avoid signal clipping in the CR processed image. CT radiation profile measurements were taken as part of routine quality control on 14 CT scanners from four vendors. CR cassettes were placed on the CT scanner bed, raised to isocenter, and leveled. Axial scans were taken at all available collimations, advancing the cassette for each scan. The CR plates were processed and raw CR data were analyzed using MATLAB scripts to measure collimation widths. The raw data approach was compared with previously established methodology. The quality control analysis scripts are released as open source using creative commons licensing. A log‐linear relationship was found between raw pixel value and air kerma, and raw data collimation width measurements were in agreement with CR‐processed, bit‐reduced data, using previously described methodology. The raw data approach, with intrinsically wider dynamic range, allows improved measurement flexibility and precision. As a result, we demonstrate a methodology for CT collimation width measurements using a single CT scan and without the need for CR scanning parameter adjustments which is more convenient for routine quality control work. PACS numbers: 87.57.Q‐, 87.59.bd, 87.57.uq John Wiley and Sons Inc. 2015-11-08 /pmc/articles/PMC5690993/ /pubmed/26699559 http://dx.doi.org/10.1120/jacmp.v16i6.5512 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Bjarnason, Thorarin A
Yang, Chang‐Ying Joseph
CT radiation profile width measurement using CR imaging plate raw data
title CT radiation profile width measurement using CR imaging plate raw data
title_full CT radiation profile width measurement using CR imaging plate raw data
title_fullStr CT radiation profile width measurement using CR imaging plate raw data
title_full_unstemmed CT radiation profile width measurement using CR imaging plate raw data
title_short CT radiation profile width measurement using CR imaging plate raw data
title_sort ct radiation profile width measurement using cr imaging plate raw data
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690993/
https://www.ncbi.nlm.nih.gov/pubmed/26699559
http://dx.doi.org/10.1120/jacmp.v16i6.5512
work_keys_str_mv AT bjarnasonthorarina ctradiationprofilewidthmeasurementusingcrimagingplaterawdata
AT yangchangyingjoseph ctradiationprofilewidthmeasurementusingcrimagingplaterawdata