Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam

This study explores the potential of cone‐beam computed tomography (CBCT) for monitoring relative beam range variations due to daily changes in patient anatomy for prostate treatment by anterior proton beams. CBCT was used to image an anthropomorphic pelvic phantom, in eight sessions on eight differ...

Descripción completa

Detalles Bibliográficos
Autores principales: Bentefour, El Hassane, Both, Stefan, Tang, Shikui, Lu, Hsiao‐Ming
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691006/
https://www.ncbi.nlm.nih.gov/pubmed/26699545
http://dx.doi.org/10.1120/jacmp.v16i6.5212
_version_ 1783279705010470912
author Bentefour, El Hassane
Both, Stefan
Tang, Shikui
Lu, Hsiao‐Ming
author_facet Bentefour, El Hassane
Both, Stefan
Tang, Shikui
Lu, Hsiao‐Ming
author_sort Bentefour, El Hassane
collection PubMed
description This study explores the potential of cone‐beam computed tomography (CBCT) for monitoring relative beam range variations due to daily changes in patient anatomy for prostate treatment by anterior proton beams. CBCT was used to image an anthropomorphic pelvic phantom, in eight sessions on eight different days. In each session, the phantom was scanned twice, first at a standard position as determined by the room lasers, and then after it was shifted by 10 mm translation randomly along one of the X, Y, or Z directions. The filling of the phantom bladder with water was not refreshed from day to day, inducing gradual change of the water‐equivalent path length (WEPL) across the bladder. MIMvista (MIM) software was used to perform image registration and re‐alignment of all the scans with the scan from the first session. The XiO treatment planning system was used to perform data analysis. It was found that, although the Hounsfield unit numbers in CBCT have substantially larger fluctuations than those in diagnostic CT, CBCT datasets taken for daily patient positioning could potentially be used to monitor changes in patient anatomy. The reproducibility of the WEPL, computed using CBCT along anterior–posterior (AP) paths across and around the phantom prostate, over a volume of 360 cc, is sufficient for detecting daily WEPL variations that are equal to or larger than 3 mm. This result also applies to CBCT scans of the phantom after it is randomly shifted from the treatment position by 10 mm. limiting the interest to WEPL variation over a specific path within the same CBCT slice, one can detect WEPL variation smaller than 1 mm. That is the case when using CBCT for tracking daily change of the WEPL across the phantom bladder that was induced by spontaneous change in the bladder filling due to evaporation. In summary, the phantom study suggests that CBCT can be used for monitoring day to day WEPL variations in a patient. The method can detect WEPL variation equal to or greater than 3 mm. The study calls for further investigation using the CBCT data from real patients. If confirmed with real patients' data, CBCT could become, in addition to patient setup, a standard tool for proton therapy pretreatment beam range check. PACS number: 87.55.Tm
format Online
Article
Text
id pubmed-5691006
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56910062018-04-02 Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam Bentefour, El Hassane Both, Stefan Tang, Shikui Lu, Hsiao‐Ming J Appl Clin Med Phys Technical Notes This study explores the potential of cone‐beam computed tomography (CBCT) for monitoring relative beam range variations due to daily changes in patient anatomy for prostate treatment by anterior proton beams. CBCT was used to image an anthropomorphic pelvic phantom, in eight sessions on eight different days. In each session, the phantom was scanned twice, first at a standard position as determined by the room lasers, and then after it was shifted by 10 mm translation randomly along one of the X, Y, or Z directions. The filling of the phantom bladder with water was not refreshed from day to day, inducing gradual change of the water‐equivalent path length (WEPL) across the bladder. MIMvista (MIM) software was used to perform image registration and re‐alignment of all the scans with the scan from the first session. The XiO treatment planning system was used to perform data analysis. It was found that, although the Hounsfield unit numbers in CBCT have substantially larger fluctuations than those in diagnostic CT, CBCT datasets taken for daily patient positioning could potentially be used to monitor changes in patient anatomy. The reproducibility of the WEPL, computed using CBCT along anterior–posterior (AP) paths across and around the phantom prostate, over a volume of 360 cc, is sufficient for detecting daily WEPL variations that are equal to or larger than 3 mm. This result also applies to CBCT scans of the phantom after it is randomly shifted from the treatment position by 10 mm. limiting the interest to WEPL variation over a specific path within the same CBCT slice, one can detect WEPL variation smaller than 1 mm. That is the case when using CBCT for tracking daily change of the WEPL across the phantom bladder that was induced by spontaneous change in the bladder filling due to evaporation. In summary, the phantom study suggests that CBCT can be used for monitoring day to day WEPL variations in a patient. The method can detect WEPL variation equal to or greater than 3 mm. The study calls for further investigation using the CBCT data from real patients. If confirmed with real patients' data, CBCT could become, in addition to patient setup, a standard tool for proton therapy pretreatment beam range check. PACS number: 87.55.Tm John Wiley and Sons Inc. 2015-11-08 /pmc/articles/PMC5691006/ /pubmed/26699545 http://dx.doi.org/10.1120/jacmp.v16i6.5212 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Bentefour, El Hassane
Both, Stefan
Tang, Shikui
Lu, Hsiao‐Ming
Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
title Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
title_full Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
title_fullStr Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
title_full_unstemmed Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
title_short Using CBCT for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
title_sort using cbct for pretreatment range check in proton therapy: a phantom study for prostate treatment by anterior–posterior beam
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691006/
https://www.ncbi.nlm.nih.gov/pubmed/26699545
http://dx.doi.org/10.1120/jacmp.v16i6.5212
work_keys_str_mv AT bentefourelhassane usingcbctforpretreatmentrangecheckinprotontherapyaphantomstudyforprostatetreatmentbyanteriorposteriorbeam
AT bothstefan usingcbctforpretreatmentrangecheckinprotontherapyaphantomstudyforprostatetreatmentbyanteriorposteriorbeam
AT tangshikui usingcbctforpretreatmentrangecheckinprotontherapyaphantomstudyforprostatetreatmentbyanteriorposteriorbeam
AT luhsiaoming usingcbctforpretreatmentrangecheckinprotontherapyaphantomstudyforprostatetreatmentbyanteriorposteriorbeam