Calibration coefficients of dosimeters used in mammography for various target/filter combinations

A wide variety of existing combinations of target and filter materials (target/filter combinations) are used in mammography equipment. The patient dose depends on the X‐ray quality that is derived from the target/filter combination, and a calibration of the dosimeter that is used in the measurement...

Descripción completa

Detalles Bibliográficos
Autores principales: Sekimoto, Michiharu, Katoh, Yoh, Katoh, Tsuguhisa
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691010/
https://www.ncbi.nlm.nih.gov/pubmed/26699570
http://dx.doi.org/10.1120/jacmp.v16i6.5604
_version_ 1783279706135592960
author Sekimoto, Michiharu
Katoh, Yoh
Katoh, Tsuguhisa
author_facet Sekimoto, Michiharu
Katoh, Yoh
Katoh, Tsuguhisa
author_sort Sekimoto, Michiharu
collection PubMed
description A wide variety of existing combinations of target and filter materials (target/filter combinations) are used in mammography equipment. The patient dose depends on the X‐ray quality that is derived from the target/filter combination, and a calibration of the dosimeter that is used in the measurement that corresponds to the specific target/filter combination is necessary. However, dosimeters in mammography are generally calibrated with reference to the X‐ray quality of Mo/Mo or W/Al combinations, and it is unclear whether or not the X‐ray quality that is derived from other target/filter combinations will affect the calibration coefficients. In this paper, the calibration coefficients of different dosimeters were evaluated for target/filter combinations. For an ionization chamber‐type dosimeter, good energy dependence was found and the effect of the target/filter combination was small. However, for a semiconductor dosimeter, a large energy dependence was found, and different calibration coefficients that depended on the target/filter combination used were required. PACS number: 87.59.E‐
format Online
Article
Text
id pubmed-5691010
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56910102018-04-02 Calibration coefficients of dosimeters used in mammography for various target/filter combinations Sekimoto, Michiharu Katoh, Yoh Katoh, Tsuguhisa J Appl Clin Med Phys Medical Imaging A wide variety of existing combinations of target and filter materials (target/filter combinations) are used in mammography equipment. The patient dose depends on the X‐ray quality that is derived from the target/filter combination, and a calibration of the dosimeter that is used in the measurement that corresponds to the specific target/filter combination is necessary. However, dosimeters in mammography are generally calibrated with reference to the X‐ray quality of Mo/Mo or W/Al combinations, and it is unclear whether or not the X‐ray quality that is derived from other target/filter combinations will affect the calibration coefficients. In this paper, the calibration coefficients of different dosimeters were evaluated for target/filter combinations. For an ionization chamber‐type dosimeter, good energy dependence was found and the effect of the target/filter combination was small. However, for a semiconductor dosimeter, a large energy dependence was found, and different calibration coefficients that depended on the target/filter combination used were required. PACS number: 87.59.E‐ John Wiley and Sons Inc. 2015-11-08 /pmc/articles/PMC5691010/ /pubmed/26699570 http://dx.doi.org/10.1120/jacmp.v16i6.5604 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Sekimoto, Michiharu
Katoh, Yoh
Katoh, Tsuguhisa
Calibration coefficients of dosimeters used in mammography for various target/filter combinations
title Calibration coefficients of dosimeters used in mammography for various target/filter combinations
title_full Calibration coefficients of dosimeters used in mammography for various target/filter combinations
title_fullStr Calibration coefficients of dosimeters used in mammography for various target/filter combinations
title_full_unstemmed Calibration coefficients of dosimeters used in mammography for various target/filter combinations
title_short Calibration coefficients of dosimeters used in mammography for various target/filter combinations
title_sort calibration coefficients of dosimeters used in mammography for various target/filter combinations
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691010/
https://www.ncbi.nlm.nih.gov/pubmed/26699570
http://dx.doi.org/10.1120/jacmp.v16i6.5604
work_keys_str_mv AT sekimotomichiharu calibrationcoefficientsofdosimetersusedinmammographyforvarioustargetfiltercombinations
AT katohyoh calibrationcoefficientsofdosimetersusedinmammographyforvarioustargetfiltercombinations
AT katohtsuguhisa calibrationcoefficientsofdosimetersusedinmammographyforvarioustargetfiltercombinations