Practical dose delivery verification of craniospinal IMRT

Craniospinal irradiation (CSI) using IMRT allows dose sparing to organs‐at‐risk (OAR) whilst conforming the dose to the target volume. Due to the complexity of treatment involving several isocenters, the dose distribution created by the inverse‐planned segmented fields must be verified prior to trea...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Young K, Kim, Anthony T, Zhao, Peiying, Karotki, Aliaksandr
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691021/
https://www.ncbi.nlm.nih.gov/pubmed/26699557
http://dx.doi.org/10.1120/jacmp.v16i6.5481
_version_ 1783279708778004480
author Lee, Young K
Kim, Anthony T
Zhao, Peiying
Karotki, Aliaksandr
author_facet Lee, Young K
Kim, Anthony T
Zhao, Peiying
Karotki, Aliaksandr
author_sort Lee, Young K
collection PubMed
description Craniospinal irradiation (CSI) using IMRT allows dose sparing to organs‐at‐risk (OAR) whilst conforming the dose to the target volume. Due to the complexity of treatment involving several isocenters, the dose distribution created by the inverse‐planned segmented fields must be verified prior to treatment. We propose and test methods to verify dose delivery using commonly available dosimetry equipment for commissioning and routine patient‐specific dose verification of craniospinal IMRT. Ten patients treated with conventional CSI were retrospectively planned with a 3‐isocenter (cranial, upper spine, and lower spine) IMRT technique. The isocenters were placed 25–27 cm away from each other in the longitudinal direction but in the same lateral and anterioposterior positions. The planning target volume (PTV) was defined as the brain with a 0.5 cm margin and spinal canal with a 1.0 cm margin. The plans were prescribed to 36 Gy in 20 fractions to the PTV mean dose. Eleven beams (five cranial, three upper spine, and three lower spine) were optimized simultaneously. The dose delivered by the IMRT plans was then recalculated on several different phantoms and measured using the following methods: 1) ionization chamber inserted in a cylindrical phantom, positioned in the junction regions between cranial/upper‐spine isocenters and upper‐/lower‐spine isocenters; 2) MapCHECK centered in the overlap regions; and 3) ArcCHECK measurement with beams from each isocenter. For 1) [Formula: see text] dose difference and for 2) and 3) [Formula: see text] of measured points with a γ‐index [Formula: see text] for 3% dose difference and 3 mm distance‐to‐agreement were deemed clinically acceptable. The median (range) planned to measured dose differences for the ionization chamber is 0.4% ([Formula: see text] to 3.0%) for the cranial/upper‐spine field and 1.8% ([Formula: see text] to 2.6%) for the upper‐/lower‐spine field overlap region. The median (range) percentage of MapCHECK diodes with a γ index [Formula: see text] for 3%/3 mm criterion is 98.0% (95.3% to 99.7%) for the cranial/upper‐spine and 97.3% (95.0% to 99.6%) for the upper‐/lower‐spine field overlap regions. The median (range) percentage of ArcCHECK diodes with a γ index [Formula: see text] for 3%/3 mm criterion is 99.7% (97.1% to 100%). Three different methods of verifying craniospinal IMRT were compared and tested. All techniques offer different benefits and together can be used for 1) commissioning the treatment technique and, separately, for 2) patient‐specific pretreatment verification measurements. PACS number: 87.55.km
format Online
Article
Text
id pubmed-5691021
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-56910212018-04-02 Practical dose delivery verification of craniospinal IMRT Lee, Young K Kim, Anthony T Zhao, Peiying Karotki, Aliaksandr J Appl Clin Med Phys Radiation Oncology Physics Craniospinal irradiation (CSI) using IMRT allows dose sparing to organs‐at‐risk (OAR) whilst conforming the dose to the target volume. Due to the complexity of treatment involving several isocenters, the dose distribution created by the inverse‐planned segmented fields must be verified prior to treatment. We propose and test methods to verify dose delivery using commonly available dosimetry equipment for commissioning and routine patient‐specific dose verification of craniospinal IMRT. Ten patients treated with conventional CSI were retrospectively planned with a 3‐isocenter (cranial, upper spine, and lower spine) IMRT technique. The isocenters were placed 25–27 cm away from each other in the longitudinal direction but in the same lateral and anterioposterior positions. The planning target volume (PTV) was defined as the brain with a 0.5 cm margin and spinal canal with a 1.0 cm margin. The plans were prescribed to 36 Gy in 20 fractions to the PTV mean dose. Eleven beams (five cranial, three upper spine, and three lower spine) were optimized simultaneously. The dose delivered by the IMRT plans was then recalculated on several different phantoms and measured using the following methods: 1) ionization chamber inserted in a cylindrical phantom, positioned in the junction regions between cranial/upper‐spine isocenters and upper‐/lower‐spine isocenters; 2) MapCHECK centered in the overlap regions; and 3) ArcCHECK measurement with beams from each isocenter. For 1) [Formula: see text] dose difference and for 2) and 3) [Formula: see text] of measured points with a γ‐index [Formula: see text] for 3% dose difference and 3 mm distance‐to‐agreement were deemed clinically acceptable. The median (range) planned to measured dose differences for the ionization chamber is 0.4% ([Formula: see text] to 3.0%) for the cranial/upper‐spine field and 1.8% ([Formula: see text] to 2.6%) for the upper‐/lower‐spine field overlap region. The median (range) percentage of MapCHECK diodes with a γ index [Formula: see text] for 3%/3 mm criterion is 98.0% (95.3% to 99.7%) for the cranial/upper‐spine and 97.3% (95.0% to 99.6%) for the upper‐/lower‐spine field overlap regions. The median (range) percentage of ArcCHECK diodes with a γ index [Formula: see text] for 3%/3 mm criterion is 99.7% (97.1% to 100%). Three different methods of verifying craniospinal IMRT were compared and tested. All techniques offer different benefits and together can be used for 1) commissioning the treatment technique and, separately, for 2) patient‐specific pretreatment verification measurements. PACS number: 87.55.km John Wiley and Sons Inc. 2015-11-08 /pmc/articles/PMC5691021/ /pubmed/26699557 http://dx.doi.org/10.1120/jacmp.v16i6.5481 Text en © 2015 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Lee, Young K
Kim, Anthony T
Zhao, Peiying
Karotki, Aliaksandr
Practical dose delivery verification of craniospinal IMRT
title Practical dose delivery verification of craniospinal IMRT
title_full Practical dose delivery verification of craniospinal IMRT
title_fullStr Practical dose delivery verification of craniospinal IMRT
title_full_unstemmed Practical dose delivery verification of craniospinal IMRT
title_short Practical dose delivery verification of craniospinal IMRT
title_sort practical dose delivery verification of craniospinal imrt
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691021/
https://www.ncbi.nlm.nih.gov/pubmed/26699557
http://dx.doi.org/10.1120/jacmp.v16i6.5481
work_keys_str_mv AT leeyoungk practicaldosedeliveryverificationofcraniospinalimrt
AT kimanthonyt practicaldosedeliveryverificationofcraniospinalimrt
AT zhaopeiying practicaldosedeliveryverificationofcraniospinalimrt
AT karotkialiaksandr practicaldosedeliveryverificationofcraniospinalimrt