Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25(th) May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroductio...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5701776/ https://www.ncbi.nlm.nih.gov/pubmed/29187798 http://dx.doi.org/10.11613/BM.2018.010705 |
_version_ | 1783281403861925888 |
---|---|
author | Koerbin, Gus Liu, Jiakai Eigenstetter, Alex Tan, Chin Hon Badrick, Tony Loh, Tze Ping |
author_facet | Koerbin, Gus Liu, Jiakai Eigenstetter, Alex Tan, Chin Hon Badrick, Tony Loh, Tze Ping |
author_sort | Koerbin, Gus |
collection | PubMed |
description | INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25(th) May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12(th) September 2016, a second reagent recall was made on 20(th) March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor. MATERIALS AND METHODS: The practices of an index laboratory were examined to determine how the analytical shifts evaded detection by routine internal quality control (IQC) and external quality assurance (EQA) systems. The ability of the patient result-based approaches, including moving average (MovAvg) and moving sum of outliers (MovSO) approaches in detecting these shifts were examined. RESULTS: Internal quality control data of the index laboratory were acceptable prior to the product recall. The practice of adjusting IQC target following a change in assay method resulted in the missed negative shift when the revised Roche assay was reintroduced. While the EQA data of the Roche subgroup showed clear negative bias relative to other laboratory methods, the results were considered as possible ‘matrix effect’. The MovAvg method detected the positive shift before the product recall. The MovSO did not detect the negative shift in the index laboratory but did so in another laboratory 5 days before the second product recall. CONCLUSIONS: There are gaps in current laboratory quality practices that leave room for analytical errors to evade detection. |
format | Online Article Text |
id | pubmed-5701776 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-57017762017-11-29 Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices Koerbin, Gus Liu, Jiakai Eigenstetter, Alex Tan, Chin Hon Badrick, Tony Loh, Tze Ping Biochem Med (Zagreb) Original Papers INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25(th) May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12(th) September 2016, a second reagent recall was made on 20(th) March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor. MATERIALS AND METHODS: The practices of an index laboratory were examined to determine how the analytical shifts evaded detection by routine internal quality control (IQC) and external quality assurance (EQA) systems. The ability of the patient result-based approaches, including moving average (MovAvg) and moving sum of outliers (MovSO) approaches in detecting these shifts were examined. RESULTS: Internal quality control data of the index laboratory were acceptable prior to the product recall. The practice of adjusting IQC target following a change in assay method resulted in the missed negative shift when the revised Roche assay was reintroduced. While the EQA data of the Roche subgroup showed clear negative bias relative to other laboratory methods, the results were considered as possible ‘matrix effect’. The MovAvg method detected the positive shift before the product recall. The MovSO did not detect the negative shift in the index laboratory but did so in another laboratory 5 days before the second product recall. CONCLUSIONS: There are gaps in current laboratory quality practices that leave room for analytical errors to evade detection. Croatian Society of Medical Biochemistry and Laboratory Medicine 2017-11-24 2018-02-15 /pmc/articles/PMC5701776/ /pubmed/29187798 http://dx.doi.org/10.11613/BM.2018.010705 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Papers Koerbin, Gus Liu, Jiakai Eigenstetter, Alex Tan, Chin Hon Badrick, Tony Loh, Tze Ping Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
title | Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
title_full | Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
title_fullStr | Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
title_full_unstemmed | Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
title_short | Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
title_sort | missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices |
topic | Original Papers |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5701776/ https://www.ncbi.nlm.nih.gov/pubmed/29187798 http://dx.doi.org/10.11613/BM.2018.010705 |
work_keys_str_mv | AT koerbingus misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices AT liujiakai misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices AT eigenstetteralex misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices AT tanchinhon misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices AT badricktony misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices AT lohtzeping misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices |