Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices

INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25(th) May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroductio...

Descripción completa

Detalles Bibliográficos
Autores principales: Koerbin, Gus, Liu, Jiakai, Eigenstetter, Alex, Tan, Chin Hon, Badrick, Tony, Loh, Tze Ping
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5701776/
https://www.ncbi.nlm.nih.gov/pubmed/29187798
http://dx.doi.org/10.11613/BM.2018.010705
_version_ 1783281403861925888
author Koerbin, Gus
Liu, Jiakai
Eigenstetter, Alex
Tan, Chin Hon
Badrick, Tony
Loh, Tze Ping
author_facet Koerbin, Gus
Liu, Jiakai
Eigenstetter, Alex
Tan, Chin Hon
Badrick, Tony
Loh, Tze Ping
author_sort Koerbin, Gus
collection PubMed
description INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25(th) May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12(th) September 2016, a second reagent recall was made on 20(th) March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor. MATERIALS AND METHODS: The practices of an index laboratory were examined to determine how the analytical shifts evaded detection by routine internal quality control (IQC) and external quality assurance (EQA) systems. The ability of the patient result-based approaches, including moving average (MovAvg) and moving sum of outliers (MovSO) approaches in detecting these shifts were examined. RESULTS: Internal quality control data of the index laboratory were acceptable prior to the product recall. The practice of adjusting IQC target following a change in assay method resulted in the missed negative shift when the revised Roche assay was reintroduced. While the EQA data of the Roche subgroup showed clear negative bias relative to other laboratory methods, the results were considered as possible ‘matrix effect’. The MovAvg method detected the positive shift before the product recall. The MovSO did not detect the negative shift in the index laboratory but did so in another laboratory 5 days before the second product recall. CONCLUSIONS: There are gaps in current laboratory quality practices that leave room for analytical errors to evade detection.
format Online
Article
Text
id pubmed-5701776
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-57017762017-11-29 Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices Koerbin, Gus Liu, Jiakai Eigenstetter, Alex Tan, Chin Hon Badrick, Tony Loh, Tze Ping Biochem Med (Zagreb) Original Papers INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25(th) May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12(th) September 2016, a second reagent recall was made on 20(th) March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor. MATERIALS AND METHODS: The practices of an index laboratory were examined to determine how the analytical shifts evaded detection by routine internal quality control (IQC) and external quality assurance (EQA) systems. The ability of the patient result-based approaches, including moving average (MovAvg) and moving sum of outliers (MovSO) approaches in detecting these shifts were examined. RESULTS: Internal quality control data of the index laboratory were acceptable prior to the product recall. The practice of adjusting IQC target following a change in assay method resulted in the missed negative shift when the revised Roche assay was reintroduced. While the EQA data of the Roche subgroup showed clear negative bias relative to other laboratory methods, the results were considered as possible ‘matrix effect’. The MovAvg method detected the positive shift before the product recall. The MovSO did not detect the negative shift in the index laboratory but did so in another laboratory 5 days before the second product recall. CONCLUSIONS: There are gaps in current laboratory quality practices that leave room for analytical errors to evade detection. Croatian Society of Medical Biochemistry and Laboratory Medicine 2017-11-24 2018-02-15 /pmc/articles/PMC5701776/ /pubmed/29187798 http://dx.doi.org/10.11613/BM.2018.010705 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Papers
Koerbin, Gus
Liu, Jiakai
Eigenstetter, Alex
Tan, Chin Hon
Badrick, Tony
Loh, Tze Ping
Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
title Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
title_full Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
title_fullStr Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
title_full_unstemmed Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
title_short Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
title_sort missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
topic Original Papers
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5701776/
https://www.ncbi.nlm.nih.gov/pubmed/29187798
http://dx.doi.org/10.11613/BM.2018.010705
work_keys_str_mv AT koerbingus misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices
AT liujiakai misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices
AT eigenstetteralex misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices
AT tanchinhon misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices
AT badricktony misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices
AT lohtzeping misseddetectionofsignificantpositiveandnegativeshiftsingentamicinassayimplicationsforroutinelaboratoryqualitypractices