Cargando…

Validation of whole room indirect calorimeters: refinement of current methodologies

Whole room indirect calorimeter (WRIC) validation techniques consist of propane combustion (PC) or infusion of mixed carbon dioxide (CO (2)) and nitrogen (N(2)) by a precision blender (PB). To determine the best method, PC of 6, 10, 22‐h and PB infusions of 6, 10, and 14‐h, were conducted. The 14‐h...

Descripción completa

Detalles Bibliográficos
Autores principales: Rising, Russell, Foerster, Thomas, Arad, Avigdor D., Albu, Jeanine, Pi‐Sunyer, Xavier
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5704087/
https://www.ncbi.nlm.nih.gov/pubmed/29180485
http://dx.doi.org/10.14814/phy2.13521
Descripción
Sumario:Whole room indirect calorimeter (WRIC) validation techniques consist of propane combustion (PC) or infusion of mixed carbon dioxide (CO (2)) and nitrogen (N(2)) by a precision blender (PB). To determine the best method, PC of 6, 10, 22‐h and PB infusions of 6, 10, and 14‐h, were conducted. The 14‐h infusion consisted of two metabolic settings. Energy expenditure (EE; kJ), ventilation (V; liters/min) of oxygen (VO (2)), VCO (2), and respiratory quotient (VCO (2)/VO (2)) obtained from the WRIC were extrapolated to the respective test durations and compared to similarly calculated values. Moreover, accurate equations (AE) were derived to correct infusions for additional N(2). As a final evaluation of a PC validated WRIC, weight maintenance (WM), energy balance (EB), respiratory quotient (RQ), and food quotients (FQ) were determined in 22 subjects who had repeat 24‐h EE measurements. Statistical analyses (P < 0.05) were conducted (SPSS, version 23). Significant differences in RQ existed between PC and stoichiometry after 6‐h. Errors for the rest of the PC tests ranged from −1.5 ± 2.4 (VCO (2)) to 2.8 ± 4.6% (EE). When compared with the WRIC, all uncorrected metabolic parameters for six and 10‐h PB infusions were significantly different with errors from −12.8 ± 1.6 (VO (2)) to 6.0 ± 2.8% (RQ). The AE reduced the magnitude of errors to −12.4 ± 1.5 (RQ) to 2.2 ± 3.0% (RQ). The PB infusion with two settings showed similar performance. No differences in WM, EB, RQ, or FQ existed in the subjects. In conclusion, 10‐h PC tests are sufficient for validating WRICs.