A respiratory compensating system: design and performance evaluation

This study proposes a respiratory compensating system which is mounted on the top of the treatment couch for reverse motion, opposite from the direction of the targets (diaphragm and hemostatic clip), in order to offset organ displacement generated by respiratory motion. Traditionally, in the treatm...

Descripción completa

Detalles Bibliográficos
Autores principales: Chuang, Ho‐Chiao, Huang, Ding‐Yang, Tien, Der‐Chi, Wu, Ren‐Hong, Hsu, Chung‐Hsien
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5711063/
https://www.ncbi.nlm.nih.gov/pubmed/24892345
http://dx.doi.org/10.1120/jacmp.v15i3.4710
_version_ 1783283000158453760
author Chuang, Ho‐Chiao
Huang, Ding‐Yang
Tien, Der‐Chi
Wu, Ren‐Hong
Hsu, Chung‐Hsien
author_facet Chuang, Ho‐Chiao
Huang, Ding‐Yang
Tien, Der‐Chi
Wu, Ren‐Hong
Hsu, Chung‐Hsien
author_sort Chuang, Ho‐Chiao
collection PubMed
description This study proposes a respiratory compensating system which is mounted on the top of the treatment couch for reverse motion, opposite from the direction of the targets (diaphragm and hemostatic clip), in order to offset organ displacement generated by respiratory motion. Traditionally, in the treatment of cancer patients, doctors must increase the field size for radiation therapy of tumors because organs move with respiratory motion, which causes radiation‐induced inflammation on the normal tissues (organ at risk (OAR)) while killing cancer cells, and thereby reducing the patient's quality of life. This study uses a strain gauge as a respiratory signal capture device to obtain abdomen respiratory signals, a proposed respiratory simulation system (RSS) and respiratory compensating system to experiment how to offset the organ displacement caused by respiratory movement and compensation effect. This study verifies the effect of the respiratory compensating system in offsetting the target displacement using two methods. The first method uses linac (medical linear accelerator) to irradiate a 300 cGy dose on the EBT film (GAFCHROMIC EBT film). The second method uses a strain gauge to capture the patients' respiratory signals, while using fluoroscopy to observe in vivo targets, such as a diaphragm, to enable the respiratory compensating system to offset the displacements of targets in superior‐inferior (SI) direction. Testing results show that the RSS position error is approximately 0.45 ~ 1.42 mm, while the respiratory compensating system position error is approximately 0.48 ~ 1.42 mm. From the EBT film profiles based on different input to the RSS, the results suggest that when the input respiratory signals of RSS are sine wave signals, the average dose (%) in the target area is improved by 1.4% ~ 24.4%, and improved in the 95% isodose area by 15.3% ~ 76.9% after compensation. If the respiratory signals input into the RSS respiratory signals are actual human respiratory signals, the average dose (%) in the target area is improved by 31.8% ~ 67.7%, and improved in the 95% isodose area by 15.3% ~ 86.4% (the above rates of improvements will increase with increasing respiratory motion displacement) after compensation. The experimental results from the second method suggested that about 67.3% ~ 82.5% displacement can be offset. In addition, gamma passing rate after compensation can be improved to 100% only when the displacement of the respiratory motion is within 10 ~ 30 mm. This study proves that the proposed system can contribute to the compensation of organ displacement caused by respiratory motion, enabling physicians to use lower doses and smaller field sizes in the treatment of tumors of cancer patients. PACS number: 87.19. Wx; 87.55. Km
format Online
Article
Text
id pubmed-5711063
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57110632018-04-02 A respiratory compensating system: design and performance evaluation Chuang, Ho‐Chiao Huang, Ding‐Yang Tien, Der‐Chi Wu, Ren‐Hong Hsu, Chung‐Hsien J Appl Clin Med Phys Radiation Protection & Regulations This study proposes a respiratory compensating system which is mounted on the top of the treatment couch for reverse motion, opposite from the direction of the targets (diaphragm and hemostatic clip), in order to offset organ displacement generated by respiratory motion. Traditionally, in the treatment of cancer patients, doctors must increase the field size for radiation therapy of tumors because organs move with respiratory motion, which causes radiation‐induced inflammation on the normal tissues (organ at risk (OAR)) while killing cancer cells, and thereby reducing the patient's quality of life. This study uses a strain gauge as a respiratory signal capture device to obtain abdomen respiratory signals, a proposed respiratory simulation system (RSS) and respiratory compensating system to experiment how to offset the organ displacement caused by respiratory movement and compensation effect. This study verifies the effect of the respiratory compensating system in offsetting the target displacement using two methods. The first method uses linac (medical linear accelerator) to irradiate a 300 cGy dose on the EBT film (GAFCHROMIC EBT film). The second method uses a strain gauge to capture the patients' respiratory signals, while using fluoroscopy to observe in vivo targets, such as a diaphragm, to enable the respiratory compensating system to offset the displacements of targets in superior‐inferior (SI) direction. Testing results show that the RSS position error is approximately 0.45 ~ 1.42 mm, while the respiratory compensating system position error is approximately 0.48 ~ 1.42 mm. From the EBT film profiles based on different input to the RSS, the results suggest that when the input respiratory signals of RSS are sine wave signals, the average dose (%) in the target area is improved by 1.4% ~ 24.4%, and improved in the 95% isodose area by 15.3% ~ 76.9% after compensation. If the respiratory signals input into the RSS respiratory signals are actual human respiratory signals, the average dose (%) in the target area is improved by 31.8% ~ 67.7%, and improved in the 95% isodose area by 15.3% ~ 86.4% (the above rates of improvements will increase with increasing respiratory motion displacement) after compensation. The experimental results from the second method suggested that about 67.3% ~ 82.5% displacement can be offset. In addition, gamma passing rate after compensation can be improved to 100% only when the displacement of the respiratory motion is within 10 ~ 30 mm. This study proves that the proposed system can contribute to the compensation of organ displacement caused by respiratory motion, enabling physicians to use lower doses and smaller field sizes in the treatment of tumors of cancer patients. PACS number: 87.19. Wx; 87.55. Km John Wiley and Sons Inc. 2014-05-08 /pmc/articles/PMC5711063/ /pubmed/24892345 http://dx.doi.org/10.1120/jacmp.v15i3.4710 Text en © 2014 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Protection & Regulations
Chuang, Ho‐Chiao
Huang, Ding‐Yang
Tien, Der‐Chi
Wu, Ren‐Hong
Hsu, Chung‐Hsien
A respiratory compensating system: design and performance evaluation
title A respiratory compensating system: design and performance evaluation
title_full A respiratory compensating system: design and performance evaluation
title_fullStr A respiratory compensating system: design and performance evaluation
title_full_unstemmed A respiratory compensating system: design and performance evaluation
title_short A respiratory compensating system: design and performance evaluation
title_sort respiratory compensating system: design and performance evaluation
topic Radiation Protection & Regulations
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5711063/
https://www.ncbi.nlm.nih.gov/pubmed/24892345
http://dx.doi.org/10.1120/jacmp.v15i3.4710
work_keys_str_mv AT chuanghochiao arespiratorycompensatingsystemdesignandperformanceevaluation
AT huangdingyang arespiratorycompensatingsystemdesignandperformanceevaluation
AT tienderchi arespiratorycompensatingsystemdesignandperformanceevaluation
AT wurenhong arespiratorycompensatingsystemdesignandperformanceevaluation
AT hsuchunghsien arespiratorycompensatingsystemdesignandperformanceevaluation
AT chuanghochiao respiratorycompensatingsystemdesignandperformanceevaluation
AT huangdingyang respiratorycompensatingsystemdesignandperformanceevaluation
AT tienderchi respiratorycompensatingsystemdesignandperformanceevaluation
AT wurenhong respiratorycompensatingsystemdesignandperformanceevaluation
AT hsuchunghsien respiratorycompensatingsystemdesignandperformanceevaluation