Data documenting the performance of the PT/INR line correction method for reconciling INR discrepancies between central laboratory coagulation analyzers using different thromboplastins during the evaluation of a portable Coagulometer
The data presented here was produced as part of an evaluation of the performance of the CoaguChek XS point-of-care coagulation analyzer, which is discussed in the research article “POCT PT INR – Is it adequate for Patient Care? A Comparison of the Roche Coaguchek XS vs. Stago Star vs. Siemens BCS in...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5711667/ https://www.ncbi.nlm.nih.gov/pubmed/29204476 http://dx.doi.org/10.1016/j.dib.2017.11.020 |
Sumario: | The data presented here was produced as part of an evaluation of the performance of the CoaguChek XS point-of-care coagulation analyzer, which is discussed in the research article “POCT PT INR – Is it adequate for Patient Care? A Comparison of the Roche Coaguchek XS vs. Stago Star vs. Siemens BCS in Patients Routinely Seen in an Anticoagulation Clinic” (Baker et al., in press) [1]. An effort to reconcile discrepancies in the patient INR result distributions from different central lab instruments (Stago Star and Siemens BCS) with the PT/INR line method is described (Poller et al., 2010, 2011; Ibrahim et al., 2011) [2], [3], [4]. While regression analysis of the ECAA Poller calibrant data provided a linear PT/INR line for all methods, Pearson's chi-squared and one-way repeated measures ANOVA analyses showed that central lab INR measurements continued to exhibit measurement site dependence after the PT/INR line correction was applied. According to paired t-test analysis, only the human thromboplastin dependent methods (CoaguChek XS and Siemens BCS both before and after PT/INR line correction) showed statistically significant agreement (p-value >0.05). |
---|