Cargando…

Fatigue fracture of a cemented Omnifit CoCr femoral stem: implant and failure analysis

A cemented, cast CoCr alloy, Omnifit Plus femoral stem was retrieved following mid-stem fracture after 24 years in vivo. The patient was an active 55-year-old male with a high body mass index (31.3) and no traumatic incidents before stem fracture. Fractographic and fatigue-based failure analyses wer...

Descripción completa

Detalles Bibliográficos
Autores principales: Bonnheim, Noah, Gramling, Hannah, Ries, Michael, Shukla, Sanjai, Iliescu, Beatrice, Pruitt, Lisa
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5712028/
https://www.ncbi.nlm.nih.gov/pubmed/29204487
http://dx.doi.org/10.1016/j.artd.2017.06.005
Descripción
Sumario:A cemented, cast CoCr alloy, Omnifit Plus femoral stem was retrieved following mid-stem fracture after 24 years in vivo. The patient was an active 55-year-old male with a high body mass index (31.3) and no traumatic incidents before stem fracture. Fractographic and fatigue-based failure analyses were performed to illuminate the etiology of fracture and retrospectively predict the device lifetime. The fracture surfaces show evidence of a coarse grain microstructure, intergranular fracture, and regions of porosity. The failure analysis suggests that stems with similar metallurgical characteristics, biomechanical environments, and in vivo durations may be abutting their functioning lifetimes, raising the possibility of an increased revision burden.