Cargando…

Bead debonding from a modern cementless total hip femoral component with concomitant taper corrosion

We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary total hi...

Descripción completa

Detalles Bibliográficos
Autores principales: Edelstein, Adam I., Phan, Duy L., Sporer, Scott M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5712029/
https://www.ncbi.nlm.nih.gov/pubmed/29204483
http://dx.doi.org/10.1016/j.artd.2017.04.003
Descripción
Sumario:We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary total hip arthroplasty, and radiographs revealed significant osteolysis and metallic debris around the femoral stem. Intraoperatively, the patient had a grossly loose femoral component with debonding of sintered beads from the femoral stem, as well as evidence of taper corrosion. We identify a failure of a modern beaded femoral component in conjunction with taper corrosion.