Corrosion and adverse local tissue reaction after total hip arthroplasty with a modular titanium alloy femoral neck
This report describes a case of mechanically assisted crevice corrosion and secondary adverse local tissue reaction in a patient following a total hip arthroplasty, utilizing a modular neck (bi-modular) femoral component. Radiographic evaluation demonstrated a well-positioned, stable, cementless art...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5712034/ https://www.ncbi.nlm.nih.gov/pubmed/29204482 http://dx.doi.org/10.1016/j.artd.2017.03.003 |
Sumario: | This report describes a case of mechanically assisted crevice corrosion and secondary adverse local tissue reaction in a patient following a total hip arthroplasty, utilizing a modular neck (bi-modular) femoral component. Radiographic evaluation demonstrated a well-positioned, stable, cementless arthroplasty. Upon further evaluation, the patient had elevated serum cobalt and chromium levels, and magnetic resonance imaging demonstrated a periprosthetic pseudotumor. Corrosion of both the neck-stem and head-neck junctions was suspected. At the time of surgery, the neck-body junction was pristine; however, the head-neck junction of the implant demonstrated severe corrosive wear, a problem that has been reported only once previously with this particular bi-modular implant. This serves as a reminder that any modular junction may be susceptible to corrosion and not all bi-modular designs behave similarly. |
---|