Commissioning and implementation of an implantable dosimeter for radiation therapy
In this article we describe commissioning and implementation procedures for the Dose Verification System (DVS) with permanently implanted in vivo wireless, telemetric radiation dosimeters for absolute dose measurements. The dosimeter uses a semiconductor device called a metal–oxide semiconductor fie...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714364/ https://www.ncbi.nlm.nih.gov/pubmed/23470929 http://dx.doi.org/10.1120/jacmp.v14i2.3989 |
_version_ | 1783283567999057920 |
---|---|
author | Buzurovic, Ivan Showalter, Timothy N. Studenski, Matthew T. Den, Robert B. Dicker, Adam P. Cao, Junsheng Xiao, Ying Yu, Yan Harrison, Amy |
author_facet | Buzurovic, Ivan Showalter, Timothy N. Studenski, Matthew T. Den, Robert B. Dicker, Adam P. Cao, Junsheng Xiao, Ying Yu, Yan Harrison, Amy |
author_sort | Buzurovic, Ivan |
collection | PubMed |
description | In this article we describe commissioning and implementation procedures for the Dose Verification System (DVS) with permanently implanted in vivo wireless, telemetric radiation dosimeters for absolute dose measurements. The dosimeter uses a semiconductor device called a metal–oxide semiconductor field‐effect transistor (MOSFET) to measure radiation dose. A MOSFET is a transistor that is generally used for amplifying or switching electronic signals. The implantable dosimeter was implemented with the goal of verifying the dose delivered to radiation therapy patients. For the purpose of acceptance testing, commissioning, and clinical implementation and to evaluate characteristics of the dosimeter, the following tests were performed: 1) temperature dependence, 2) reproducibility, 3) field size dependence, 4) postirradiation signal drift, 5) dependence on average dose rate, 6) linearity test, 7) angular dependence (different gantry angle position), 8) angular dependence (different DVS angle position), 9) dose rate dependence, 10) irradiation depth dependence, 11) effect of cone‐beam exposure to the dosimeter, and 12) multiple reading effect. The dosimeter is not currently calibrated for use in the kV range; nonetheless, the effect of the cone‐beam procedure on the MOSFET dosimeter was investigated. Phantom studies were performed in both air and water using an Elekta Synergy S Beam‐Modulator linear accelerator. Commissioning and clinical implementation for prostate cancer patients receiving external‐beam radiation therapy were performed in compliance with the general recommendations given for in vivo dosimetry devices. The reproducibility test in water at human body temperature (37°C) showed a 1.4% absolute difference, with a standard deviation of 5.72 cGy (i.e., [Formula: see text]). The constancy test shows that the average readings at room temperature were 3% lower compared to the readings at human body temperature, with a [Formula: see text]. Measurements were not dependent upon field size. Due to postirradiation signal drift, the following corrections are suggested: [Formula: see text] , [Formula: see text] , 0.5%, and 2.5% for the readings taken after 0.5, 1, 5, or 10 min, respectively. Different gantry angles did not influence the readings. The maximum error was less than 1% with a maximum [Formula: see text] (1.8%) for the gantry angle of 45°. However, readings are dependent on the dosimeter orientation. The average dose reading was 7.89 cGy ([Formula: see text]) when CBCT imaging was used for the pelvis protocol, and when postirradiation measurement was taken at 2.5 min (expected 2–3 cGy). The clinical implementation of the implantable MOSFET dosimeters for prostate cancer radiation therapy is described. Measurements performed for commissioning show that the dosimeter, if used within specifications, provides sufficient accuracy for its intended use in clinical procedures. The postradiation signal drift, temperature dependence, variation of reproducibility, and rotational isotropy could be encountered if the dosimeter is used outside the manufacturer's specifications. The dosimeter can be used as a tool for quantifying dose at depth, as well as to evaluate adherence between planned doses and the delivered doses. Currently, the system is clinically implemented with [Formula: see text] tolerance. PACS numbers: 87.53.‐j; 87.55.‐x |
format | Online Article Text |
id | pubmed-5714364 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57143642018-04-02 Commissioning and implementation of an implantable dosimeter for radiation therapy Buzurovic, Ivan Showalter, Timothy N. Studenski, Matthew T. Den, Robert B. Dicker, Adam P. Cao, Junsheng Xiao, Ying Yu, Yan Harrison, Amy J Appl Clin Med Phys Radiation Measurements In this article we describe commissioning and implementation procedures for the Dose Verification System (DVS) with permanently implanted in vivo wireless, telemetric radiation dosimeters for absolute dose measurements. The dosimeter uses a semiconductor device called a metal–oxide semiconductor field‐effect transistor (MOSFET) to measure radiation dose. A MOSFET is a transistor that is generally used for amplifying or switching electronic signals. The implantable dosimeter was implemented with the goal of verifying the dose delivered to radiation therapy patients. For the purpose of acceptance testing, commissioning, and clinical implementation and to evaluate characteristics of the dosimeter, the following tests were performed: 1) temperature dependence, 2) reproducibility, 3) field size dependence, 4) postirradiation signal drift, 5) dependence on average dose rate, 6) linearity test, 7) angular dependence (different gantry angle position), 8) angular dependence (different DVS angle position), 9) dose rate dependence, 10) irradiation depth dependence, 11) effect of cone‐beam exposure to the dosimeter, and 12) multiple reading effect. The dosimeter is not currently calibrated for use in the kV range; nonetheless, the effect of the cone‐beam procedure on the MOSFET dosimeter was investigated. Phantom studies were performed in both air and water using an Elekta Synergy S Beam‐Modulator linear accelerator. Commissioning and clinical implementation for prostate cancer patients receiving external‐beam radiation therapy were performed in compliance with the general recommendations given for in vivo dosimetry devices. The reproducibility test in water at human body temperature (37°C) showed a 1.4% absolute difference, with a standard deviation of 5.72 cGy (i.e., [Formula: see text]). The constancy test shows that the average readings at room temperature were 3% lower compared to the readings at human body temperature, with a [Formula: see text]. Measurements were not dependent upon field size. Due to postirradiation signal drift, the following corrections are suggested: [Formula: see text] , [Formula: see text] , 0.5%, and 2.5% for the readings taken after 0.5, 1, 5, or 10 min, respectively. Different gantry angles did not influence the readings. The maximum error was less than 1% with a maximum [Formula: see text] (1.8%) for the gantry angle of 45°. However, readings are dependent on the dosimeter orientation. The average dose reading was 7.89 cGy ([Formula: see text]) when CBCT imaging was used for the pelvis protocol, and when postirradiation measurement was taken at 2.5 min (expected 2–3 cGy). The clinical implementation of the implantable MOSFET dosimeters for prostate cancer radiation therapy is described. Measurements performed for commissioning show that the dosimeter, if used within specifications, provides sufficient accuracy for its intended use in clinical procedures. The postradiation signal drift, temperature dependence, variation of reproducibility, and rotational isotropy could be encountered if the dosimeter is used outside the manufacturer's specifications. The dosimeter can be used as a tool for quantifying dose at depth, as well as to evaluate adherence between planned doses and the delivered doses. Currently, the system is clinically implemented with [Formula: see text] tolerance. PACS numbers: 87.53.‐j; 87.55.‐x John Wiley and Sons Inc. 2013-03-04 /pmc/articles/PMC5714364/ /pubmed/23470929 http://dx.doi.org/10.1120/jacmp.v14i2.3989 Text en © 2013 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Measurements Buzurovic, Ivan Showalter, Timothy N. Studenski, Matthew T. Den, Robert B. Dicker, Adam P. Cao, Junsheng Xiao, Ying Yu, Yan Harrison, Amy Commissioning and implementation of an implantable dosimeter for radiation therapy |
title | Commissioning and implementation of an implantable dosimeter for radiation therapy |
title_full | Commissioning and implementation of an implantable dosimeter for radiation therapy |
title_fullStr | Commissioning and implementation of an implantable dosimeter for radiation therapy |
title_full_unstemmed | Commissioning and implementation of an implantable dosimeter for radiation therapy |
title_short | Commissioning and implementation of an implantable dosimeter for radiation therapy |
title_sort | commissioning and implementation of an implantable dosimeter for radiation therapy |
topic | Radiation Measurements |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714364/ https://www.ncbi.nlm.nih.gov/pubmed/23470929 http://dx.doi.org/10.1120/jacmp.v14i2.3989 |
work_keys_str_mv | AT buzurovicivan commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT showaltertimothyn commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT studenskimatthewt commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT denrobertb commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT dickeradamp commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT caojunsheng commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT xiaoying commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT yuyan commissioningandimplementationofanimplantabledosimeterforradiationtherapy AT harrisonamy commissioningandimplementationofanimplantabledosimeterforradiationtherapy |