The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner

Radiochromic and radiographic films are widely used for radiation dosimetry due to the advantage of high spatial resolution and two‐dimensional dose measurement. Different types of scanners, including various models of flatbed scanners, have been used as part of the dosimetry readout procedure. This...

Descripción completa

Detalles Bibliográficos
Autores principales: Sim, GS, Wong, JHD, Ng, KH
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714532/
https://www.ncbi.nlm.nih.gov/pubmed/23835383
http://dx.doi.org/10.1120/jacmp.v14i4.4182
_version_ 1783283600116940800
author Sim, GS
Wong, JHD
Ng, KH
author_facet Sim, GS
Wong, JHD
Ng, KH
author_sort Sim, GS
collection PubMed
description Radiochromic and radiographic films are widely used for radiation dosimetry due to the advantage of high spatial resolution and two‐dimensional dose measurement. Different types of scanners, including various models of flatbed scanners, have been used as part of the dosimetry readout procedure. This paper focuses on the characterization of the EBT2 film response in combination with a Microtek ScanMaker 9800XL scanner and the subsequent use in the dosimetric verification of a 3D conformal radiotherapy treatment. The film reproducibility and scanner uniformity of the Microtek ScanMaker 9800XL was studied. A three‐field 3D conformal radiotherapy treatment was planned on an anthropomorphic phantom and EBT2 film measurements were carried out to verify the treatment. The interfilm reproducibility was found to be 0.25%. Over a period of three months, the films darkened by 1%. The scanner reproducibility was ± 2% and a nonuniformity was [Formula: see text] along the direction perpendicular to the scan direction. EBT2 measurements showed an underdose of 6.2% at high‐dose region compared to TPS predicted dose. This may be due to the inability of the treatment planning system to predict the correct dose distribution in the presence of tissue inhomogeneities and the uncertainty of the scanner reproducibility and uniformity. The use of EBT2 film in conjunction with the axial CT image of the anthropomorphic phantom allows the evaluation of the anatomical location of dose discrepancies between the EBT2 measured dose distribution and TPS predicted dose distribution. PACS number: 87.55.Qr
format Online
Article
Text
id pubmed-5714532
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57145322018-04-02 The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner Sim, GS Wong, JHD Ng, KH J Appl Clin Med Phys Radiation Oncology Physics Radiochromic and radiographic films are widely used for radiation dosimetry due to the advantage of high spatial resolution and two‐dimensional dose measurement. Different types of scanners, including various models of flatbed scanners, have been used as part of the dosimetry readout procedure. This paper focuses on the characterization of the EBT2 film response in combination with a Microtek ScanMaker 9800XL scanner and the subsequent use in the dosimetric verification of a 3D conformal radiotherapy treatment. The film reproducibility and scanner uniformity of the Microtek ScanMaker 9800XL was studied. A three‐field 3D conformal radiotherapy treatment was planned on an anthropomorphic phantom and EBT2 film measurements were carried out to verify the treatment. The interfilm reproducibility was found to be 0.25%. Over a period of three months, the films darkened by 1%. The scanner reproducibility was ± 2% and a nonuniformity was [Formula: see text] along the direction perpendicular to the scan direction. EBT2 measurements showed an underdose of 6.2% at high‐dose region compared to TPS predicted dose. This may be due to the inability of the treatment planning system to predict the correct dose distribution in the presence of tissue inhomogeneities and the uncertainty of the scanner reproducibility and uniformity. The use of EBT2 film in conjunction with the axial CT image of the anthropomorphic phantom allows the evaluation of the anatomical location of dose discrepancies between the EBT2 measured dose distribution and TPS predicted dose distribution. PACS number: 87.55.Qr John Wiley and Sons Inc. 2013-07-08 /pmc/articles/PMC5714532/ /pubmed/23835383 http://dx.doi.org/10.1120/jacmp.v14i4.4182 Text en © 2013 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Sim, GS
Wong, JHD
Ng, KH
The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
title The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
title_full The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
title_fullStr The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
title_full_unstemmed The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
title_short The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner
title_sort use of radiochromic ebt2 film for the quality assurance and dosimetric verification of 3d conformal radiotherapy using microtek scanmaker 9800xl flatbed scanner
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714532/
https://www.ncbi.nlm.nih.gov/pubmed/23835383
http://dx.doi.org/10.1120/jacmp.v14i4.4182
work_keys_str_mv AT simgs theuseofradiochromicebt2filmforthequalityassuranceanddosimetricverificationof3dconformalradiotherapyusingmicrotekscanmaker9800xlflatbedscanner
AT wongjhd theuseofradiochromicebt2filmforthequalityassuranceanddosimetricverificationof3dconformalradiotherapyusingmicrotekscanmaker9800xlflatbedscanner
AT ngkh theuseofradiochromicebt2filmforthequalityassuranceanddosimetricverificationof3dconformalradiotherapyusingmicrotekscanmaker9800xlflatbedscanner
AT simgs useofradiochromicebt2filmforthequalityassuranceanddosimetricverificationof3dconformalradiotherapyusingmicrotekscanmaker9800xlflatbedscanner
AT wongjhd useofradiochromicebt2filmforthequalityassuranceanddosimetricverificationof3dconformalradiotherapyusingmicrotekscanmaker9800xlflatbedscanner
AT ngkh useofradiochromicebt2filmforthequalityassuranceanddosimetricverificationof3dconformalradiotherapyusingmicrotekscanmaker9800xlflatbedscanner