A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac

Several investigators have pointed out that electron and neutron contamination from high‐energy photon beams are clinically important. The aim of this study is to assess electron and neutron contamination production by various prostheses in a high‐energy photon beam of a medical linac. A 15 MV Sieme...

Descripción completa

Detalles Bibliográficos
Autores principales: Toossi, Mohammad Taghi Bahreyni, Behmadi, Marziyeh, Ghorbani, Mahdi, Gholamhosseinian, Hamid
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714559/
https://www.ncbi.nlm.nih.gov/pubmed/24036859
http://dx.doi.org/10.1120/jacmp.v14i5.4253
_version_ 1783283605664956416
author Toossi, Mohammad Taghi Bahreyni
Behmadi, Marziyeh
Ghorbani, Mahdi
Gholamhosseinian, Hamid
author_facet Toossi, Mohammad Taghi Bahreyni
Behmadi, Marziyeh
Ghorbani, Mahdi
Gholamhosseinian, Hamid
author_sort Toossi, Mohammad Taghi Bahreyni
collection PubMed
description Several investigators have pointed out that electron and neutron contamination from high‐energy photon beams are clinically important. The aim of this study is to assess electron and neutron contamination production by various prostheses in a high‐energy photon beam of a medical linac. A 15 MV Siemens PRIMUS linac was simulated by MCNPX Monte Carlo (MC) code and the results of percentage depth dose (PDD) and dose profile values were compared with the measured data. Electron and neutron contaminations were calculated on the beam's central axis for Co‐Cr‐Mo, stainless steel, Ti‐alloy, and Ti hip prostheses through MC simulations. Dose increase factor (DIF) was calculated as the ratio of electron (neutron) dose at a point for [Formula: see text] field size in presence of prosthesis to that at the same point in absence of prosthesis. DIF was estimated at different depths in a water phantom. Our MC‐calculated PDD and dose profile data are in good agreement with the corresponding measured values. Maximum dose increase factor for electron contamination for Co‐Cr‐Mo, stainless steel, Ti‐alloy, and Ti prostheses were equal to 1.18, 1.16, 1.16, and 1.14, respectively. The corresponding values for neutron contamination were respectively equal to: 184.55, 137.33, 40.66, and 43.17. Titanium‐based prostheses are recommended for the orthopedic practice of hip junction replacement. When treatment planning for a patient with hip prosthesis is performed for a high‐energy photon beam, attempt should be made to ensure that the prosthesis is not exposed to primary photons. PACS numbers: 87.56.bd, 87.55.kh, 87.55.Gh
format Online
Article
Text
id pubmed-5714559
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57145592018-04-02 A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac Toossi, Mohammad Taghi Bahreyni Behmadi, Marziyeh Ghorbani, Mahdi Gholamhosseinian, Hamid J Appl Clin Med Phys Radiation Oncology Physics Several investigators have pointed out that electron and neutron contamination from high‐energy photon beams are clinically important. The aim of this study is to assess electron and neutron contamination production by various prostheses in a high‐energy photon beam of a medical linac. A 15 MV Siemens PRIMUS linac was simulated by MCNPX Monte Carlo (MC) code and the results of percentage depth dose (PDD) and dose profile values were compared with the measured data. Electron and neutron contaminations were calculated on the beam's central axis for Co‐Cr‐Mo, stainless steel, Ti‐alloy, and Ti hip prostheses through MC simulations. Dose increase factor (DIF) was calculated as the ratio of electron (neutron) dose at a point for [Formula: see text] field size in presence of prosthesis to that at the same point in absence of prosthesis. DIF was estimated at different depths in a water phantom. Our MC‐calculated PDD and dose profile data are in good agreement with the corresponding measured values. Maximum dose increase factor for electron contamination for Co‐Cr‐Mo, stainless steel, Ti‐alloy, and Ti prostheses were equal to 1.18, 1.16, 1.16, and 1.14, respectively. The corresponding values for neutron contamination were respectively equal to: 184.55, 137.33, 40.66, and 43.17. Titanium‐based prostheses are recommended for the orthopedic practice of hip junction replacement. When treatment planning for a patient with hip prosthesis is performed for a high‐energy photon beam, attempt should be made to ensure that the prosthesis is not exposed to primary photons. PACS numbers: 87.56.bd, 87.55.kh, 87.55.Gh John Wiley and Sons Inc. 2013-09-06 /pmc/articles/PMC5714559/ /pubmed/24036859 http://dx.doi.org/10.1120/jacmp.v14i5.4253 Text en © 2013 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Toossi, Mohammad Taghi Bahreyni
Behmadi, Marziyeh
Ghorbani, Mahdi
Gholamhosseinian, Hamid
A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
title A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
title_full A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
title_fullStr A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
title_full_unstemmed A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
title_short A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
title_sort monte carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 mv siemens primus linac
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714559/
https://www.ncbi.nlm.nih.gov/pubmed/24036859
http://dx.doi.org/10.1120/jacmp.v14i5.4253
work_keys_str_mv AT toossimohammadtaghibahreyni amontecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT behmadimarziyeh amontecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT ghorbanimahdi amontecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT gholamhosseinianhamid amontecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT toossimohammadtaghibahreyni montecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT behmadimarziyeh montecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT ghorbanimahdi montecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac
AT gholamhosseinianhamid montecarlostudyonelectronandneutroncontaminationcausedbythepresenceofhipprosthesisinphotonmodeofa15mvsiemensprimuslinac