Cargando…
Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications
A glass seed consisting of the [Formula: see text] ‐emitting radionuclide [Formula: see text] incorporated with radionuclide [Formula: see text] as SPECT marker is proposed for potential application in brachytherapy in order to reduce the undesirable dose to healthy adjacent organs. The aim of this...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714572/ https://www.ncbi.nlm.nih.gov/pubmed/24036862 http://dx.doi.org/10.1120/jacmp.v14i5.4302 |
_version_ | 1783283609102188544 |
---|---|
author | Hadadi, Asghar Sadeghi, Mahdi Sardari, Dariush Khanchi, Alireza Shirazi, Alireza |
author_facet | Hadadi, Asghar Sadeghi, Mahdi Sardari, Dariush Khanchi, Alireza Shirazi, Alireza |
author_sort | Hadadi, Asghar |
collection | PubMed |
description | A glass seed consisting of the [Formula: see text] ‐emitting radionuclide [Formula: see text] incorporated with radionuclide [Formula: see text] as SPECT marker is proposed for potential application in brachytherapy in order to reduce the undesirable dose to healthy adjacent organs. The aim of this work is to determine the dosimetric characteristics, as suggested in the AAPM TG‐60/TG‐149 reports, for this seed using Monte Carlo simulation. Monte Carlo codes MCNP5, EGSnrc, and FLUKA were used to calculate the absorbed dose distribution around the seed. Dosimetric parameters, such as reference absorbed dose rate, radial dose function, and one‐dimensional (1D) and two‐dimensional (2D) anisotropy functions, were obtained. The computational results from these three codes are in agreement within 5.4% difference on average. The absorbed dose rate at the reference point was estimated to be 5.01 cGy [Formula: see text] and self absorption of YAS glass seed amounted to 30.51%. The results showed that, with thermal neutron bombardment of 5 hours in a typical flux, sufficient activity for applications in brachytherapy may be achieved. With a 5 mCi initial activity, the total dose of a YAS glass seed was estimated to be 1.38 Gy at 1.0 cm from the seed center. Comparing with gamma emitting seeds, the [Formula: see text] seed could reduce undesirable doses to adjacent organs, because of the rapid dose falloff of beta ray. Because of the high [Formula: see text] value of 5.5 mm, fewer number of [Formula: see text] seeds will be required for an interstitial brachytherapy treatment using permanent implant, in comparison with other beta‐emitting seeds. The results would be helpful in the development of the radioactive implants using [Formula: see text] glass seeds for the brachytherapy treatment. PACS numbers: 87.53.Jw, 87.56.bg |
format | Online Article Text |
id | pubmed-5714572 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57145722018-04-02 Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications Hadadi, Asghar Sadeghi, Mahdi Sardari, Dariush Khanchi, Alireza Shirazi, Alireza J Appl Clin Med Phys Radiation Oncology Physics A glass seed consisting of the [Formula: see text] ‐emitting radionuclide [Formula: see text] incorporated with radionuclide [Formula: see text] as SPECT marker is proposed for potential application in brachytherapy in order to reduce the undesirable dose to healthy adjacent organs. The aim of this work is to determine the dosimetric characteristics, as suggested in the AAPM TG‐60/TG‐149 reports, for this seed using Monte Carlo simulation. Monte Carlo codes MCNP5, EGSnrc, and FLUKA were used to calculate the absorbed dose distribution around the seed. Dosimetric parameters, such as reference absorbed dose rate, radial dose function, and one‐dimensional (1D) and two‐dimensional (2D) anisotropy functions, were obtained. The computational results from these three codes are in agreement within 5.4% difference on average. The absorbed dose rate at the reference point was estimated to be 5.01 cGy [Formula: see text] and self absorption of YAS glass seed amounted to 30.51%. The results showed that, with thermal neutron bombardment of 5 hours in a typical flux, sufficient activity for applications in brachytherapy may be achieved. With a 5 mCi initial activity, the total dose of a YAS glass seed was estimated to be 1.38 Gy at 1.0 cm from the seed center. Comparing with gamma emitting seeds, the [Formula: see text] seed could reduce undesirable doses to adjacent organs, because of the rapid dose falloff of beta ray. Because of the high [Formula: see text] value of 5.5 mm, fewer number of [Formula: see text] seeds will be required for an interstitial brachytherapy treatment using permanent implant, in comparison with other beta‐emitting seeds. The results would be helpful in the development of the radioactive implants using [Formula: see text] glass seeds for the brachytherapy treatment. PACS numbers: 87.53.Jw, 87.56.bg John Wiley and Sons Inc. 2013-09-06 /pmc/articles/PMC5714572/ /pubmed/24036862 http://dx.doi.org/10.1120/jacmp.v14i5.4302 Text en © 2013 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Hadadi, Asghar Sadeghi, Mahdi Sardari, Dariush Khanchi, Alireza Shirazi, Alireza Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications |
title | Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications |
title_full | Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications |
title_fullStr | Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications |
title_full_unstemmed | Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications |
title_short | Monte Carlo characterization of biocompatible beta‐emitting [Formula: see text] glass seed incorporated with the radionuclide [Formula: see text] as a SPECT marker for brachytherapy applications |
title_sort | monte carlo characterization of biocompatible beta‐emitting [formula: see text] glass seed incorporated with the radionuclide [formula: see text] as a spect marker for brachytherapy applications |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714572/ https://www.ncbi.nlm.nih.gov/pubmed/24036862 http://dx.doi.org/10.1120/jacmp.v14i5.4302 |
work_keys_str_mv | AT hadadiasghar montecarlocharacterizationofbiocompatiblebetaemittingformulaseetextglassseedincorporatedwiththeradionuclideformulaseetextasaspectmarkerforbrachytherapyapplications AT sadeghimahdi montecarlocharacterizationofbiocompatiblebetaemittingformulaseetextglassseedincorporatedwiththeradionuclideformulaseetextasaspectmarkerforbrachytherapyapplications AT sardaridariush montecarlocharacterizationofbiocompatiblebetaemittingformulaseetextglassseedincorporatedwiththeradionuclideformulaseetextasaspectmarkerforbrachytherapyapplications AT khanchialireza montecarlocharacterizationofbiocompatiblebetaemittingformulaseetextglassseedincorporatedwiththeradionuclideformulaseetextasaspectmarkerforbrachytherapyapplications AT shirazialireza montecarlocharacterizationofbiocompatiblebetaemittingformulaseetextglassseedincorporatedwiththeradionuclideformulaseetextasaspectmarkerforbrachytherapyapplications |