Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx
Small field dosimetry is a challenging task. The difficulties of small field measurements, particularly stereotactic field size measurements, are highlighted by the large interinstitution variability that can be observed for circular cone collimator commissioning measurements. We believe the best wa...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714633/ https://www.ncbi.nlm.nih.gov/pubmed/24257290 http://dx.doi.org/10.1120/jacmp.v14i6.4493 |
_version_ | 1783283623807418368 |
---|---|
author | Wiant, David B. Terrell, Jonathon A. Maurer, Jacqueline M. Yount, Caroline L. Sintay, Benjamin J. |
author_facet | Wiant, David B. Terrell, Jonathon A. Maurer, Jacqueline M. Yount, Caroline L. Sintay, Benjamin J. |
author_sort | Wiant, David B. |
collection | PubMed |
description | Small field dosimetry is a challenging task. The difficulties of small field measurements, particularly stereotactic field size measurements, are highlighted by the large interinstitution variability that can be observed for circular cone collimator commissioning measurements. We believe the best way to improve the consistency of small field measurements is to clearly document and share the results of small field measurements. In this work we report on the commissioning and validation of a BrainLAB cone system for 6 MV and 10 MV flattening filter‐free (FFF) beams on a Varian TrueBeam STx. Commissioning measurements consisted of output factors, percent depth dose, and off‐axis factor measurements with a diode. Validation measurements were made in a polystyrene slab phantom at depths of 5 cm, 10 cm, and 15 cm using radiochromic film. Output factors for the 6xFFF cones are 0.689, 0. 790, 0.830, 0.871, 0.890, and 0.901 for 4 mm, 6 mm, 7.5 mm, 10 mm, 12.5 mm, and the 15 mm cones, respectively. Output factors for the 10xFFF cones are 0.566, 0. 699, 0.756, 0.826, 0.864, and 0.888 for 4 mm, 6 mm, 7.5 mm, 10 mm, 12.5 mm, and the 15 mm cones, respectively. The full width half maximum values of the off‐axis factors agreed with the nominal cone size to within 0.5 mm. Validation measurements showed an agreement of absolute dose between calculation and plan of ≤ 3.6%, and an agreement of field sizes of ≤ 0.3 mm in all cases. Radiochromic film validation measurements show reasonable agreement with beam models for circular collimators based on diode commissioning measurements. PACS numbers: 87.53.Ly, 87.53.Bn, 87.56.nk, 87.55.D‐, 87.55.km |
format | Online Article Text |
id | pubmed-5714633 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57146332018-04-02 Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx Wiant, David B. Terrell, Jonathon A. Maurer, Jacqueline M. Yount, Caroline L. Sintay, Benjamin J. J Appl Clin Med Phys Radiation Oncology Physics Small field dosimetry is a challenging task. The difficulties of small field measurements, particularly stereotactic field size measurements, are highlighted by the large interinstitution variability that can be observed for circular cone collimator commissioning measurements. We believe the best way to improve the consistency of small field measurements is to clearly document and share the results of small field measurements. In this work we report on the commissioning and validation of a BrainLAB cone system for 6 MV and 10 MV flattening filter‐free (FFF) beams on a Varian TrueBeam STx. Commissioning measurements consisted of output factors, percent depth dose, and off‐axis factor measurements with a diode. Validation measurements were made in a polystyrene slab phantom at depths of 5 cm, 10 cm, and 15 cm using radiochromic film. Output factors for the 6xFFF cones are 0.689, 0. 790, 0.830, 0.871, 0.890, and 0.901 for 4 mm, 6 mm, 7.5 mm, 10 mm, 12.5 mm, and the 15 mm cones, respectively. Output factors for the 10xFFF cones are 0.566, 0. 699, 0.756, 0.826, 0.864, and 0.888 for 4 mm, 6 mm, 7.5 mm, 10 mm, 12.5 mm, and the 15 mm cones, respectively. The full width half maximum values of the off‐axis factors agreed with the nominal cone size to within 0.5 mm. Validation measurements showed an agreement of absolute dose between calculation and plan of ≤ 3.6%, and an agreement of field sizes of ≤ 0.3 mm in all cases. Radiochromic film validation measurements show reasonable agreement with beam models for circular collimators based on diode commissioning measurements. PACS numbers: 87.53.Ly, 87.53.Bn, 87.56.nk, 87.55.D‐, 87.55.km John Wiley and Sons Inc. 2013-11-04 /pmc/articles/PMC5714633/ /pubmed/24257290 http://dx.doi.org/10.1120/jacmp.v14i6.4493 Text en © 2013 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Wiant, David B. Terrell, Jonathon A. Maurer, Jacqueline M. Yount, Caroline L. Sintay, Benjamin J. Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx |
title | Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx |
title_full | Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx |
title_fullStr | Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx |
title_full_unstemmed | Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx |
title_short | Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx |
title_sort | commissioning and validation of brainlab cones for 6x fff and 10x fff beams on a varian truebeam stx |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714633/ https://www.ncbi.nlm.nih.gov/pubmed/24257290 http://dx.doi.org/10.1120/jacmp.v14i6.4493 |
work_keys_str_mv | AT wiantdavidb commissioningandvalidationofbrainlabconesfor6xfffand10xfffbeamsonavariantruebeamstx AT terrelljonathona commissioningandvalidationofbrainlabconesfor6xfffand10xfffbeamsonavariantruebeamstx AT maurerjacquelinem commissioningandvalidationofbrainlabconesfor6xfffand10xfffbeamsonavariantruebeamstx AT yountcarolinel commissioningandvalidationofbrainlabconesfor6xfffand10xfffbeamsonavariantruebeamstx AT sintaybenjaminj commissioningandvalidationofbrainlabconesfor6xfffand10xfffbeamsonavariantruebeamstx |