Cargando…
A novel compound 6D‐offset simulating phantom and quality assurance program for stereotactic image‐guided radiation therapy system
A comprehensive quality assurance (QA) device cum program was developed for the commissioning and routine testing of the 6D IGRT systems. In this article, both the new QA system and the BrainLAB IGRT system which was added onto a Varian Clinac were evaluated. A novel compound 6D‐offset simulating ph...
Autores principales: | Ngar, Dennis Yuen‐Kan, Cheung, Michael Lok‐Man, Kam, Michael Koon‐Ming, Poon, Wai‐Sang, Chan, Anthony Tak‐Cheung |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714637/ https://www.ncbi.nlm.nih.gov/pubmed/24257294 http://dx.doi.org/10.1120/jacmp.v14i6.4297 |
Ejemplares similares
-
Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials
por: Faught, Austin M., et al.
Publicado: (2013) -
Teflon cylindrical phantom for delivery quality assurance of SBRT
por: Lack, Danielle W., et al.
Publicado: (2014) -
Anthropomorphic breast phantoms for quality assurance and dose verification
por: Bencomo, José A., et al.
Publicado: (2004) -
Patient‐specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT)
por: Chow, Vivian U. Y., et al.
Publicado: (2020) -
Evaluation of the RUBY modular QA phantom for planar and non‐coplanar VMAT and stereotactic radiations
por: Poppinga, Daniela, et al.
Publicado: (2020)