Modulation factors calculated with an EPID‐derived MLC fluence model to streamline IMRT/VMAT second checks
This work outlines the development of a robust method of calculating modulation factors used for the independent verification of MUs for IMRT and VMAT treatments, to replace onerous ion chamber measurements. Two‐dimensional fluence maps were calculated for dynamic MLC fields that include MLC interle...