Cargando…

Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°

Remote afterloading devices used for high‐dose‐rate (HDR) brachytherapy may be supplied with different sources, and these sources typically have differing initial source strengths. In addition, the proposed frequency for source changes may also vary, depending upon the source type. Dosimetric parame...

Descripción completa

Detalles Bibliográficos
Autores principales: Davis, Stephen D., Parker, William, Evans, Michael D.C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714643/
https://www.ncbi.nlm.nih.gov/pubmed/24257270
http://dx.doi.org/10.1120/jacmp.v14i6.4239
_version_ 1783283626116382720
author Davis, Stephen D.
Parker, William
Evans, Michael D.C.
author_facet Davis, Stephen D.
Parker, William
Evans, Michael D.C.
author_sort Davis, Stephen D.
collection PubMed
description Remote afterloading devices used for high‐dose‐rate (HDR) brachytherapy may be supplied with different sources, and these sources typically have differing initial source strengths. In addition, the proposed frequency for source changes may also vary, depending upon the source type. Dosimetric parameters unique to each source are often used to compare source types. However, when considering the relative dosimetric efficiency between two HDR sources, the combined effect of source type, initial source strength, and source change scheme must be considered. A method of quantifying this combined effect by calculating mean dose rate from specific dosimetric source data is discussed. This method suggests an objective manner of comparing source scheme equivalency to facilitate performing a cost ratio analysis between different HDR sources and source change schemes. PACS numbers: 87.53.Jw, 87.56.bg
format Online
Article
Text
id pubmed-5714643
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57146432018-04-02 Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy° Davis, Stephen D. Parker, William Evans, Michael D.C. J Appl Clin Med Phys Radiation Oncology Physics Remote afterloading devices used for high‐dose‐rate (HDR) brachytherapy may be supplied with different sources, and these sources typically have differing initial source strengths. In addition, the proposed frequency for source changes may also vary, depending upon the source type. Dosimetric parameters unique to each source are often used to compare source types. However, when considering the relative dosimetric efficiency between two HDR sources, the combined effect of source type, initial source strength, and source change scheme must be considered. A method of quantifying this combined effect by calculating mean dose rate from specific dosimetric source data is discussed. This method suggests an objective manner of comparing source scheme equivalency to facilitate performing a cost ratio analysis between different HDR sources and source change schemes. PACS numbers: 87.53.Jw, 87.56.bg John Wiley and Sons Inc. 2013-11-04 /pmc/articles/PMC5714643/ /pubmed/24257270 http://dx.doi.org/10.1120/jacmp.v14i6.4239 Text en © 2013 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Davis, Stephen D.
Parker, William
Evans, Michael D.C.
Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
title Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
title_full Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
title_fullStr Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
title_full_unstemmed Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
title_short Using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
title_sort using mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for hdr brachytherapy°
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714643/
https://www.ncbi.nlm.nih.gov/pubmed/24257270
http://dx.doi.org/10.1120/jacmp.v14i6.4239
work_keys_str_mv AT davisstephend usingmeandoseratetocomparerelativedosimetricefficiencywithrespecttosourcetypeandsourcechangeschedulesforhdrbrachytherapy
AT parkerwilliam usingmeandoseratetocomparerelativedosimetricefficiencywithrespecttosourcetypeandsourcechangeschedulesforhdrbrachytherapy
AT evansmichaeldc usingmeandoseratetocomparerelativedosimetricefficiencywithrespecttosourcetypeandsourcechangeschedulesforhdrbrachytherapy