Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix

Heterogeneities existing in the patient during treatment are neglected, as the treated subject is considered homogeneous in most of the commercially‐available treatment planning systems (TPSs) used for high dose rate (HDR) brachytherapy. The choice of a suitable dosimeter for experimental dosimetry...

Descripción completa

Detalles Bibliográficos
Autores principales: Uniyal, Satish C., Naithani, Umesh C., Sharma, Sunil D., Srivastava, Anoop K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5716133/
https://www.ncbi.nlm.nih.gov/pubmed/22231217
http://dx.doi.org/10.1120/jacmp.v13i1.3654
_version_ 1783283882687201280
author Uniyal, Satish C.
Naithani, Umesh C.
Sharma, Sunil D.
Srivastava, Anoop K.
author_facet Uniyal, Satish C.
Naithani, Umesh C.
Sharma, Sunil D.
Srivastava, Anoop K.
author_sort Uniyal, Satish C.
collection PubMed
description Heterogeneities existing in the patient during treatment are neglected, as the treated subject is considered homogeneous in most of the commercially‐available treatment planning systems (TPSs) used for high dose rate (HDR) brachytherapy. The choice of a suitable dosimeter for experimental dosimetry near the HDR source is crucial, mainly due to existence of steep dose gradients. The present work aimed to assess the effect of rectal air heterogeneity and applicator attenuation in the HDR Ir‐192 brachytherapy treatment of carcinoma uterine cervix by utilizing GAFCHROMIC EBT2 film dosimetry. The dose to rectal walls under the condition of rectal air heterogeneity was measured experimentally using EBT2 film in a rectal phantom, and the measurements were validated by the Monte Carlo (MC) simulations. The applicator attenuation was measured by EBT2 film for a commonly used stainless steel uterine tube in a homogeneous water equivalent phantom. The measured doses were compared with the TPS calculated values. In case of the air cavity, the measured dose at the closest rectal surface was 12.8% less than the TPS calculated value due to lack of back scattering, whereas at the farthest rectal surface, it was higher by 24.5% due to no attenuation. The magnitude of attenuation due to the metal applicator was measured as high as 2% when compared with the TPS calculation. The dose reduction at the nearest rectal surface due to the effect of rectal air has indicated a clinically favorable dose distribution within the rectum, whereas the shielding effect posed by the metallic applicator was found to be less significant. Mutual agreement of the measured doses with the MC calculated dose values confirmed the suitability of EBT2 film for clinical dosimetry in HDR brachytherapy. PACS numbers: 87.53.Bn, 87.53.Jw, 87.56.bg, 87.55.Qr
format Online
Article
Text
id pubmed-5716133
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57161332018-04-02 Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix Uniyal, Satish C. Naithani, Umesh C. Sharma, Sunil D. Srivastava, Anoop K. J Appl Clin Med Phys Radiation Oncology Physics Heterogeneities existing in the patient during treatment are neglected, as the treated subject is considered homogeneous in most of the commercially‐available treatment planning systems (TPSs) used for high dose rate (HDR) brachytherapy. The choice of a suitable dosimeter for experimental dosimetry near the HDR source is crucial, mainly due to existence of steep dose gradients. The present work aimed to assess the effect of rectal air heterogeneity and applicator attenuation in the HDR Ir‐192 brachytherapy treatment of carcinoma uterine cervix by utilizing GAFCHROMIC EBT2 film dosimetry. The dose to rectal walls under the condition of rectal air heterogeneity was measured experimentally using EBT2 film in a rectal phantom, and the measurements were validated by the Monte Carlo (MC) simulations. The applicator attenuation was measured by EBT2 film for a commonly used stainless steel uterine tube in a homogeneous water equivalent phantom. The measured doses were compared with the TPS calculated values. In case of the air cavity, the measured dose at the closest rectal surface was 12.8% less than the TPS calculated value due to lack of back scattering, whereas at the farthest rectal surface, it was higher by 24.5% due to no attenuation. The magnitude of attenuation due to the metal applicator was measured as high as 2% when compared with the TPS calculation. The dose reduction at the nearest rectal surface due to the effect of rectal air has indicated a clinically favorable dose distribution within the rectum, whereas the shielding effect posed by the metallic applicator was found to be less significant. Mutual agreement of the measured doses with the MC calculated dose values confirmed the suitability of EBT2 film for clinical dosimetry in HDR brachytherapy. PACS numbers: 87.53.Bn, 87.53.Jw, 87.56.bg, 87.55.Qr John Wiley and Sons Inc. 2012-01-05 /pmc/articles/PMC5716133/ /pubmed/22231217 http://dx.doi.org/10.1120/jacmp.v13i1.3654 Text en © 2012 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Uniyal, Satish C.
Naithani, Umesh C.
Sharma, Sunil D.
Srivastava, Anoop K.
Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
title Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
title_full Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
title_fullStr Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
title_full_unstemmed Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
title_short Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
title_sort radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5716133/
https://www.ncbi.nlm.nih.gov/pubmed/22231217
http://dx.doi.org/10.1120/jacmp.v13i1.3654
work_keys_str_mv AT uniyalsatishc radiochromicfilmdosimetryofrectalinhomogeneityandapplicatorattenuationinhighdoseratebrachytherapyofuterinecervix
AT naithaniumeshc radiochromicfilmdosimetryofrectalinhomogeneityandapplicatorattenuationinhighdoseratebrachytherapyofuterinecervix
AT sharmasunild radiochromicfilmdosimetryofrectalinhomogeneityandapplicatorattenuationinhighdoseratebrachytherapyofuterinecervix
AT srivastavaanoopk radiochromicfilmdosimetryofrectalinhomogeneityandapplicatorattenuationinhighdoseratebrachytherapyofuterinecervix