Cargando…

Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT

The purpose of this study was to measure and compare the contrast‐to‐noise ratio (CNR) as a function of dose for the cone‐beam CT (CBCT) produced by the imaging beam line (IBL) for the standard and an investigational imaging panel. Two Siemens Artiste linear accelerators were modified at our institu...

Descripción completa

Detalles Bibliográficos
Autor principal: Beltran, Chris
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5716139/
https://www.ncbi.nlm.nih.gov/pubmed/22231211
http://dx.doi.org/10.1120/jacmp.v13i1.3607
_version_ 1783283884017844224
author Beltran, Chris
author_facet Beltran, Chris
author_sort Beltran, Chris
collection PubMed
description The purpose of this study was to measure and compare the contrast‐to‐noise ratio (CNR) as a function of dose for the cone‐beam CT (CBCT) produced by the imaging beam line (IBL) for the standard and an investigational imaging panel. Two Siemens Artiste linear accelerators were modified at our institution such that the MV‐CBCT would operate under an investigational IBL. The imaging panel from one of the machines was replaced with an investigational imaging panel. After the modification, a set of CBCT for a large and small phantom consisting of eight tissue‐equivalent inserts was acquired for the standard imager and for the investigational imager with and without the standard copper plate. Ten dose settings for each phantom using the IBL in combination with the standard and investigational imaging panel were acquired. The CNR for each tissue‐equivalent insert was calculated. Resolution measurements in line pairs per mm (lp/mm) of the CBCT for the various imaging panel setups were made. In addition, CBCT images of two patients that were imaged with each panel configuration were displayed for a group of physicians and therapists who were asked to identify the best and worst CBCT for each patient. This was used as a qualitative judge of practical image quality. The CNR of the muscle insert for the large phantom with 1.5 cGy at isocenter was 1.3 for the standard imager, 1.5 for the investigational imager with the copper plate, and 1.9 without the plate. Under the same conditions, the CNR of the trabecular bone insert was 5.9, 7.3, and 9.7, respectively. For the small phantom with the same dose to isocenter, the CNR for muscle was 1.7, 2.1, and 3.3, respectively. For the trabecular bone, the CNR was 8.1, 9.6, and 12.1 respectively. The resolution for 1 cGy at isocenter was 0.37 lp/mm for the standard imager, 0.32 and 0.33 for the investigational imager with and without the copper plate. The qualitative test ranked the CBCT of the investigational imager without the copper plate to be the best image, and the standard imager to be the worst. The investigational imaging panel improves image quality as compared to the standard imager for IBL CBCTs. A 1 cGy IBL CBCT, no matter which imager is used, is sufficient for bony anatomy localization. The investigational imager without the copper plate was judged clinically to produce the best IBL CBCT. PACS numbers: 87.57.Q, 87.57.cj
format Online
Article
Text
id pubmed-5716139
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57161392018-04-02 Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT Beltran, Chris J Appl Clin Med Phys Radiation Oncology Physics The purpose of this study was to measure and compare the contrast‐to‐noise ratio (CNR) as a function of dose for the cone‐beam CT (CBCT) produced by the imaging beam line (IBL) for the standard and an investigational imaging panel. Two Siemens Artiste linear accelerators were modified at our institution such that the MV‐CBCT would operate under an investigational IBL. The imaging panel from one of the machines was replaced with an investigational imaging panel. After the modification, a set of CBCT for a large and small phantom consisting of eight tissue‐equivalent inserts was acquired for the standard imager and for the investigational imager with and without the standard copper plate. Ten dose settings for each phantom using the IBL in combination with the standard and investigational imaging panel were acquired. The CNR for each tissue‐equivalent insert was calculated. Resolution measurements in line pairs per mm (lp/mm) of the CBCT for the various imaging panel setups were made. In addition, CBCT images of two patients that were imaged with each panel configuration were displayed for a group of physicians and therapists who were asked to identify the best and worst CBCT for each patient. This was used as a qualitative judge of practical image quality. The CNR of the muscle insert for the large phantom with 1.5 cGy at isocenter was 1.3 for the standard imager, 1.5 for the investigational imager with the copper plate, and 1.9 without the plate. Under the same conditions, the CNR of the trabecular bone insert was 5.9, 7.3, and 9.7, respectively. For the small phantom with the same dose to isocenter, the CNR for muscle was 1.7, 2.1, and 3.3, respectively. For the trabecular bone, the CNR was 8.1, 9.6, and 12.1 respectively. The resolution for 1 cGy at isocenter was 0.37 lp/mm for the standard imager, 0.32 and 0.33 for the investigational imager with and without the copper plate. The qualitative test ranked the CBCT of the investigational imager without the copper plate to be the best image, and the standard imager to be the worst. The investigational imaging panel improves image quality as compared to the standard imager for IBL CBCTs. A 1 cGy IBL CBCT, no matter which imager is used, is sufficient for bony anatomy localization. The investigational imager without the copper plate was judged clinically to produce the best IBL CBCT. PACS numbers: 87.57.Q, 87.57.cj John Wiley and Sons Inc. 2012-01-05 /pmc/articles/PMC5716139/ /pubmed/22231211 http://dx.doi.org/10.1120/jacmp.v13i1.3607 Text en © 2012 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Beltran, Chris
Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT
title Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT
title_full Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT
title_fullStr Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT
title_full_unstemmed Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT
title_short Image quality of an investigational imaging panel for use with the imaging beam line cone‐beam CT
title_sort image quality of an investigational imaging panel for use with the imaging beam line cone‐beam ct
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5716139/
https://www.ncbi.nlm.nih.gov/pubmed/22231211
http://dx.doi.org/10.1120/jacmp.v13i1.3607
work_keys_str_mv AT beltranchris imagequalityofaninvestigationalimagingpanelforusewiththeimagingbeamlineconebeamct