The Radiological Physics Center's standard dataset for small field size output factors

Delivery of accurate intensity‐modulated radiation therapy (IMRT) or stereotactic radiotherapy depends on a multitude of steps in the treatment delivery process. These steps range from imaging of the patient to dose calculation to machine delivery of the treatment plan. Within the treatment planning...

Descripción completa

Detalles Bibliográficos
Autores principales: Followill, David S., Kry, Stephen F., Qin, Lihong, Leif, Jessica, Molineu, Andrea, Alvarez, Paola, Aguirre, Jose Francisco, Ibbott, Geoffrey S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718235/
https://www.ncbi.nlm.nih.gov/pubmed/22955664
http://dx.doi.org/10.1120/jacmp.v13i5.3962
Descripción
Sumario:Delivery of accurate intensity‐modulated radiation therapy (IMRT) or stereotactic radiotherapy depends on a multitude of steps in the treatment delivery process. These steps range from imaging of the patient to dose calculation to machine delivery of the treatment plan. Within the treatment planning system's (TPS) dose calculation algorithm, various unique small field dosimetry parameters are essential, such as multileaf collimator modeling and field size dependence of the output. One of the largest challenges in this process is determining accurate small field size output factors. The Radiological Physics Center (RPC), as part of its mission to ensure that institutions deliver comparable and consistent radiation doses to their patients, conducts on‐site dosimetry review visits to institutions. As a part of the on‐site audit, the RPC measures the small field size output factors as might be used in IMRT treatments, and compares the resulting field size dependent output factors to values calculated by the institution's treatment planning system (TPS). The RPC has gathered multiple small field size output factor datasets for X‐ray energies ranging from 6 to 18 MV from Varian, Siemens and Elekta linear accelerators. These datasets were measured at 10 cm depth and ranged from [Formula: see text] to [Formula: see text]. The field sizes were defined by the MLC and for the Varian machines the secondary jaws were maintained at a [Formula: see text]. The RPC measurements were made with a micro‐ion chamber whose volume was small enough to gather a full ionization reading even for the [Formula: see text] field size. The RPC measured output factors are tabulated and are reproducible with standard deviations (SD) ranging from 0.1% to 2.4%, while the institutions' calculated values had a much larger SD range, ranging up to 7.9%. The absolute average percent differences were greater for the [Formula: see text] than for the other field sizes. The RPC's measured small field output factors provide institutions with a standard dataset against which to compare their TPS calculated values. Any discrepancies noted between the standard dataset and calculated values should be investigated with careful measurements and with attention to the specific beam model. PACS number: 87.53.Bn