Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers

The purpose of this study was to evaluate the visibility and artifact created by gold, carbon, and polymer fiducial markers in a simple phantom across computed tomography (CT), kilovoltage (kV), and megavoltage (MV) linear accelerator imaging and MV tomotherapy imaging. Three types of fiducial marke...

Descripción completa

Detalles Bibliográficos
Autores principales: Handsfield, Lydia L., Yue, Ning J., Zhou, Jinghao, Chen, Ting, Goyal, Sharad
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718239/
https://www.ncbi.nlm.nih.gov/pubmed/22955665
http://dx.doi.org/10.1120/jacmp.v13i5.3976
_version_ 1783284297666396160
author Handsfield, Lydia L.
Yue, Ning J.
Zhou, Jinghao
Chen, Ting
Goyal, Sharad
author_facet Handsfield, Lydia L.
Yue, Ning J.
Zhou, Jinghao
Chen, Ting
Goyal, Sharad
author_sort Handsfield, Lydia L.
collection PubMed
description The purpose of this study was to evaluate the visibility and artifact created by gold, carbon, and polymer fiducial markers in a simple phantom across computed tomography (CT), kilovoltage (kV), and megavoltage (MV) linear accelerator imaging and MV tomotherapy imaging. Three types of fiducial markers (gold, carbon, and polymer) were investigated for their visibility and artifacts in images acquired with various modalities and with different imaging parameters (kV, mAs, slice thickness). The imaging modalities include kV CT, 2D linac‐based kilovoltage and megavoltage X‐ray imaging systems, kV cone‐beam CT, and normal and fine tomotherapy imaging. The images were acquired on a phantom constructed using Superflab bolus in which markers of each type were inserted into the center layer. The visibility and artifacts produced by each marker were assessed qualitatively and quantitatively. All tested markers could be identified clearly on the acquired CT and linac‐based kV images; gold markers demonstrated the highest contrast. On the CT images, gold markers produced a significant artifact, while no artifacts were observed with polymer markers. Only gold markers were visible when using linac‐based MV and tomotherapy imaging. For linac‐based kV images, the contrast increased with kV and mAs values for all the markers, with the gold being the most pronounced. On CT images, the contrast increased with kV for the gold markers, while decreasing for the polymer and carbon marker. With the bolus phantom used, we found that when kV imaging‐based treatment verification equipment is available, polymer and carbon markers may be the preferred choice for target localization and patient treatment positioning verification due to less image artifacts. If MV imaging will be the sole modality for positioning verification, it may be necessary to use gold markers despite the artifacts they create on the simulation CT images. PACS number: 87
format Online
Article
Text
id pubmed-5718239
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57182392018-04-02 Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers Handsfield, Lydia L. Yue, Ning J. Zhou, Jinghao Chen, Ting Goyal, Sharad J Appl Clin Med Phys Radiation Oncology Physics The purpose of this study was to evaluate the visibility and artifact created by gold, carbon, and polymer fiducial markers in a simple phantom across computed tomography (CT), kilovoltage (kV), and megavoltage (MV) linear accelerator imaging and MV tomotherapy imaging. Three types of fiducial markers (gold, carbon, and polymer) were investigated for their visibility and artifacts in images acquired with various modalities and with different imaging parameters (kV, mAs, slice thickness). The imaging modalities include kV CT, 2D linac‐based kilovoltage and megavoltage X‐ray imaging systems, kV cone‐beam CT, and normal and fine tomotherapy imaging. The images were acquired on a phantom constructed using Superflab bolus in which markers of each type were inserted into the center layer. The visibility and artifacts produced by each marker were assessed qualitatively and quantitatively. All tested markers could be identified clearly on the acquired CT and linac‐based kV images; gold markers demonstrated the highest contrast. On the CT images, gold markers produced a significant artifact, while no artifacts were observed with polymer markers. Only gold markers were visible when using linac‐based MV and tomotherapy imaging. For linac‐based kV images, the contrast increased with kV and mAs values for all the markers, with the gold being the most pronounced. On CT images, the contrast increased with kV for the gold markers, while decreasing for the polymer and carbon marker. With the bolus phantom used, we found that when kV imaging‐based treatment verification equipment is available, polymer and carbon markers may be the preferred choice for target localization and patient treatment positioning verification due to less image artifacts. If MV imaging will be the sole modality for positioning verification, it may be necessary to use gold markers despite the artifacts they create on the simulation CT images. PACS number: 87 John Wiley and Sons Inc. 2012-09-06 /pmc/articles/PMC5718239/ /pubmed/22955665 http://dx.doi.org/10.1120/jacmp.v13i5.3976 Text en © 2012 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Handsfield, Lydia L.
Yue, Ning J.
Zhou, Jinghao
Chen, Ting
Goyal, Sharad
Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
title Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
title_full Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
title_fullStr Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
title_full_unstemmed Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
title_short Determination of optimal fiducial marker across image‐guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
title_sort determination of optimal fiducial marker across image‐guided radiation therapy (igrt) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718239/
https://www.ncbi.nlm.nih.gov/pubmed/22955665
http://dx.doi.org/10.1120/jacmp.v13i5.3976
work_keys_str_mv AT handsfieldlydial determinationofoptimalfiducialmarkeracrossimageguidedradiationtherapyigrtmodalitiesvisibilityandartifactanalysisofgoldcarbonandpolymerfiducialmarkers
AT yueningj determinationofoptimalfiducialmarkeracrossimageguidedradiationtherapyigrtmodalitiesvisibilityandartifactanalysisofgoldcarbonandpolymerfiducialmarkers
AT zhoujinghao determinationofoptimalfiducialmarkeracrossimageguidedradiationtherapyigrtmodalitiesvisibilityandartifactanalysisofgoldcarbonandpolymerfiducialmarkers
AT chenting determinationofoptimalfiducialmarkeracrossimageguidedradiationtherapyigrtmodalitiesvisibilityandartifactanalysisofgoldcarbonandpolymerfiducialmarkers
AT goyalsharad determinationofoptimalfiducialmarkeracrossimageguidedradiationtherapyigrtmodalitiesvisibilityandartifactanalysisofgoldcarbonandpolymerfiducialmarkers