Cargando…

Dose distribution changes with shielding disc misalignments and wrong orientations in breast IOERT: a Monte Carlo – GEANT4 and experimental study

One of the most relevant risks in breast intraoperative electron radiotherapy (IOERT) is the incorrect positioning of the shielding disc. If such a setup error occurs, the treatment zone could receive a nonuniform dose delivery, and a considerable part of the electron beam could hit — and irradiate...

Descripción completa

Detalles Bibliográficos
Autores principales: Russo, Giorgio, Casarino, Carlo, Arnetta, Gaetano, Candiano, Giuliana, Stefano, Alessandro, Alongi, Filippo, Borasi, Giovanni, Messa, Cristina, Gilardi, Maria C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718242/
https://www.ncbi.nlm.nih.gov/pubmed/22955646
http://dx.doi.org/10.1120/jacmp.v13i5.3817