Cargando…

A quality assurance method with submillimeter accuracy for stereotactic linear accelerators

The Stereotactic Alignment for Linear Accelerator (S. A. Linac) system is developed to conveniently improve the alignment accuracy of a conventional linac equipped with stereotactic cones. From the Winston‐Lutz test, the SAlinac system performs three‐dimensional (3D) reconstruction of the quality as...

Descripción completa

Detalles Bibliográficos
Autores principales: Grimm, Jimm, Grimm, Shu‐Ya Lisa, Das, Indra J., Zhu, Yunping, Yeo, Inhwan, Xue, Jinyu, Simpson, Larry, Jacob, Dayee, Sarkar, Abhirup
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718578/
https://www.ncbi.nlm.nih.gov/pubmed/21330986
http://dx.doi.org/10.1120/jacmp.v12i1.3365
Descripción
Sumario:The Stereotactic Alignment for Linear Accelerator (S. A. Linac) system is developed to conveniently improve the alignment accuracy of a conventional linac equipped with stereotactic cones. From the Winston‐Lutz test, the SAlinac system performs three‐dimensional (3D) reconstruction of the quality assurance (QA) ball coordinates with respect to the radiation isocenter, and combines this information with digital images of the laser target to determine the absolute position of the room lasers. A handheld device provides near‐real‐time repositioning advice to enable the user to align the QA ball and room lasers to within 0.25 mm of the centroid of the radiation isocenter. The results of 37 Winston‐Lutz tests over 68 days showed that the median 3D QA ball alignment error was 0.09 mm, and 97% of the time the 3D error was [Formula: see text]. All 3D isocentric errors in the study were 0.3 mm or less. The median x and y laser alignment coordinate error was 0.09 mm, and 94% of the time the x and y laser error was [Formula: see text]. A phantom test showed that the system can make submillimeter end‐to‐end accuracy achievable, making a conventional linac a “Submillimeter Knife”. PACS numbers: 87.53.Ly, 87.55.Qr