Use of reduced dose rate when treating moving tumors using dynamic IMRT
The purpose was to evaluate the effect of dose rate on discrepancies between expected and delivered dose caused by the interplay effect. Fifteen separate dynamic IMRT plans and five hybrid IMRT plans were created for five patients (three IMRT plans and one hybrid IMRT plan per patient). The impact o...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718590/ https://www.ncbi.nlm.nih.gov/pubmed/21330973 http://dx.doi.org/10.1120/jacmp.v12i1.3276 |
Sumario: | The purpose was to evaluate the effect of dose rate on discrepancies between expected and delivered dose caused by the interplay effect. Fifteen separate dynamic IMRT plans and five hybrid IMRT plans were created for five patients (three IMRT plans and one hybrid IMRT plan per patient). The impact of motion on the delivered dose was evaluated experimentally for each treatment field for different dose rates (200 and [Formula: see text]), and for a range of target amplitudes and periods. The maximum dose discrepancy for dynamic IMRT fields was 18.5% and 10.3% for dose rates of 400 and [Formula: see text] , respectively. The maximum dose discrepancy was larger than this for hybrid plans, but the results were similar when weighted by the contribution of the IMRT fields. The percentage of fields for which 98% of the target never experienced a 5% or 10% dose discrepancy increased when the dose rate was reduced from [Formula: see text] to [Formula: see text]. For amplitudes up to 2 cm, reducing the dose rate to [Formula: see text] is effective in keeping daily dose discrepancies for each field within 10%. PACS number: 87.55Qr |
---|