Cargando…

Evaluation of tomotherapy MVCT image enhancement program for tumor volume delineation

The aims of this study were to investigate the variability between physicians in delineation of head and neck tumors on original tomotherapy megavoltage CT (MVCT) studies and corresponding software enhanced MVCT images, and to establish an optimal approach for evaluation of image improvement. Five p...

Descripción completa

Detalles Bibliográficos
Autores principales: Martin, Spencer, Rodrigues, George, Chen, Quan, Pavamani, Simon, Read, Nancy, Ahmad, Belal, Hammond, J. Alex, Venkatesan, Varagur, Renaud, James, Yartsev, Slav
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718637/
https://www.ncbi.nlm.nih.gov/pubmed/21844864
http://dx.doi.org/10.1120/jacmp.v12i3.3505
Descripción
Sumario:The aims of this study were to investigate the variability between physicians in delineation of head and neck tumors on original tomotherapy megavoltage CT (MVCT) studies and corresponding software enhanced MVCT images, and to establish an optimal approach for evaluation of image improvement. Five physicians contoured the gross tumor volume (GTV) for three head and neck cancer patients on 34 original and enhanced MVCT studies. Variation between original and enhanced MVCT studies was quantified by DICE coefficient and the coefficient of variance. Based on volume of agreement between physicians, higher correlation in terms of average DICE coefficients was observed in GTV delineation for enhanced MVCT for patients 1, 2, and 3 by 15%, 3%, and 7%, respectively, while delineation variance among physicians was reduced using enhanced MVCT for 12 of 17 weekly image studies. Enhanced MVCT provides advantages in reduction of variance among physicians in delineation of the GTV. Agreement on contouring by the same physician on both original and enhanced MVCT was equally high. PACS numbers: 87.57.N‐, 87.57.np, 87.57.nt