Detection of patient setup errors with a portal image – DRR registration software application

The purpose of this study was to evaluate a custom portal image — digitally reconstructed radiograph (DRR) registration software application. The software works by transforming the portal image into the coordinate space of the DRR image using three control points placed on each image by the user, an...

Descripción completa

Detalles Bibliográficos
Autores principales: Sutherland, Kenneth, Ishikawa, Masayori, Bengua, Gerard, Ito, Yoichi M., Miyamoto, Yoshiko, Shirato, Hiroki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718652/
https://www.ncbi.nlm.nih.gov/pubmed/21844862
http://dx.doi.org/10.1120/jacmp.v12i3.3492
_version_ 1783284356225171456
author Sutherland, Kenneth
Ishikawa, Masayori
Bengua, Gerard
Ito, Yoichi M.
Miyamoto, Yoshiko
Shirato, Hiroki
author_facet Sutherland, Kenneth
Ishikawa, Masayori
Bengua, Gerard
Ito, Yoichi M.
Miyamoto, Yoshiko
Shirato, Hiroki
author_sort Sutherland, Kenneth
collection PubMed
description The purpose of this study was to evaluate a custom portal image — digitally reconstructed radiograph (DRR) registration software application. The software works by transforming the portal image into the coordinate space of the DRR image using three control points placed on each image by the user, and displaying the fused image. In order to test statistically that the software actually improves setup error estimation, an intra‐ and interobserver phantom study was performed. Portal images of anthropomorphic thoracic and pelvis phantoms with virtually placed irradiation fields at known setup errors were prepared. A group of five doctors was first asked to estimate the setup errors by examining the portal and DRR image side‐by‐side, not using the software. A second group of four technicians then estimated the same set of images using the registration software. These two groups of human subjects were then compared with an auto‐registration feature of the software, which is based on the mutual information between the portal and DRR images. For the thoracic case, the average distance between the actual setup error and the estimated error was [Formula: see text] for doctors using the side‐by‐side method, [Formula: see text] for technicians using the registration method, and [Formula: see text] for the automatic algorithm. For the pelvis case, the average distance between the actual setup error and estimated error was [Formula: see text] for the doctors using the side‐by‐side method, [Formula: see text] for technicians using the registration method, and [Formula: see text] for the automatic algorithm. The ability of humans to estimate offset values improved statistically using our software for the chest phantom that we tested. Setup error estimation was further improved using our automatic error estimation algorithm. Estimations were not statistically different for the pelvis case. Consistency improved using the software for both the chest and pelvis phantoms. We also tested the automatic algorithm with a database of over 5,000 clinical cases from our hospital. The algorithm performed well for head and breast but performed poorly for pelvis cases, probably due to lack of contrast in the megavoltage portal image. The software incorporates an original algorithm to fuse portal and DRR images, which we describe in detail. The offset optimization algorithm used in the automatic mode of operation is also unique, and may be useful if the contrast of the portal images can be improved. PACS numbers: 87.55.Qr, 87.57.nj
format Online
Article
Text
id pubmed-5718652
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57186522018-04-02 Detection of patient setup errors with a portal image – DRR registration software application Sutherland, Kenneth Ishikawa, Masayori Bengua, Gerard Ito, Yoichi M. Miyamoto, Yoshiko Shirato, Hiroki J Appl Clin Med Phys Radiation Oncology Physics The purpose of this study was to evaluate a custom portal image — digitally reconstructed radiograph (DRR) registration software application. The software works by transforming the portal image into the coordinate space of the DRR image using three control points placed on each image by the user, and displaying the fused image. In order to test statistically that the software actually improves setup error estimation, an intra‐ and interobserver phantom study was performed. Portal images of anthropomorphic thoracic and pelvis phantoms with virtually placed irradiation fields at known setup errors were prepared. A group of five doctors was first asked to estimate the setup errors by examining the portal and DRR image side‐by‐side, not using the software. A second group of four technicians then estimated the same set of images using the registration software. These two groups of human subjects were then compared with an auto‐registration feature of the software, which is based on the mutual information between the portal and DRR images. For the thoracic case, the average distance between the actual setup error and the estimated error was [Formula: see text] for doctors using the side‐by‐side method, [Formula: see text] for technicians using the registration method, and [Formula: see text] for the automatic algorithm. For the pelvis case, the average distance between the actual setup error and estimated error was [Formula: see text] for the doctors using the side‐by‐side method, [Formula: see text] for technicians using the registration method, and [Formula: see text] for the automatic algorithm. The ability of humans to estimate offset values improved statistically using our software for the chest phantom that we tested. Setup error estimation was further improved using our automatic error estimation algorithm. Estimations were not statistically different for the pelvis case. Consistency improved using the software for both the chest and pelvis phantoms. We also tested the automatic algorithm with a database of over 5,000 clinical cases from our hospital. The algorithm performed well for head and breast but performed poorly for pelvis cases, probably due to lack of contrast in the megavoltage portal image. The software incorporates an original algorithm to fuse portal and DRR images, which we describe in detail. The offset optimization algorithm used in the automatic mode of operation is also unique, and may be useful if the contrast of the portal images can be improved. PACS numbers: 87.55.Qr, 87.57.nj John Wiley and Sons Inc. 2011-02-18 /pmc/articles/PMC5718652/ /pubmed/21844862 http://dx.doi.org/10.1120/jacmp.v12i3.3492 Text en © 2011 The Authors. https://creativecommons.org/licenses/by/3.0/This is an open access article under the terms of the Creative Commons Attribution (https://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Sutherland, Kenneth
Ishikawa, Masayori
Bengua, Gerard
Ito, Yoichi M.
Miyamoto, Yoshiko
Shirato, Hiroki
Detection of patient setup errors with a portal image – DRR registration software application
title Detection of patient setup errors with a portal image – DRR registration software application
title_full Detection of patient setup errors with a portal image – DRR registration software application
title_fullStr Detection of patient setup errors with a portal image – DRR registration software application
title_full_unstemmed Detection of patient setup errors with a portal image – DRR registration software application
title_short Detection of patient setup errors with a portal image – DRR registration software application
title_sort detection of patient setup errors with a portal image – drr registration software application
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718652/
https://www.ncbi.nlm.nih.gov/pubmed/21844862
http://dx.doi.org/10.1120/jacmp.v12i3.3492
work_keys_str_mv AT sutherlandkenneth detectionofpatientsetuperrorswithaportalimagedrrregistrationsoftwareapplication
AT ishikawamasayori detectionofpatientsetuperrorswithaportalimagedrrregistrationsoftwareapplication
AT benguagerard detectionofpatientsetuperrorswithaportalimagedrrregistrationsoftwareapplication
AT itoyoichim detectionofpatientsetuperrorswithaportalimagedrrregistrationsoftwareapplication
AT miyamotoyoshiko detectionofpatientsetuperrorswithaportalimagedrrregistrationsoftwareapplication
AT shiratohiroki detectionofpatientsetuperrorswithaportalimagedrrregistrationsoftwareapplication