Measurement of Knee Rotation Angles Using a Smartphone Application: An Experimental Study of Porcine Knees

PURPOSE: This study evaluated the efficacy of a smartphone application in the measurement of rotation angles in porcine knees. MATERIALS AND METHODS: Two K-wires were fixated to the femoral condyle and anterior tibial crest of 10 porcine legs. The angle created between the K-wires with an external r...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Hee-June, Lee, Hyun-Joo, Shin, Ji-Yeon, Choi, Young-Seo, Kyung, Hee-Soo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Knee Society 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5718789/
https://www.ncbi.nlm.nih.gov/pubmed/29172391
http://dx.doi.org/10.5792/ksrr.17.046
Descripción
Sumario:PURPOSE: This study evaluated the efficacy of a smartphone application in the measurement of rotation angles in porcine knees. MATERIALS AND METHODS: Two K-wires were fixated to the femoral condyle and anterior tibial crest of 10 porcine legs. The angle created between the K-wires with an external rotation force applied was measured on a photograph and defined as the true angle. The same force was applied to the legs placed on a splint with a smartphone attached to the plantar side. The angle presented on a smartphone application was determined as the measured angle. The differences between the true and measured angles in 30° and 90° knee flexion and differences in measured angles depending on the status of the popliteus tendon were compared. RESULTS: In the intact knees, the mean true angles in 30° and 90° flexion were 20.5°±1.4° and 19.1°±1.3°, respectively, and the mean measured angles in 30° and 90° flexion were 21.1°±0.9° and 18.6°±1.6°, respectively. When the popliteus tendon was cut, the mean true angles in 30° and 90° flexion were 31.4°±1.1° and 38.5°±2.5°, respectively, and the mean measured angles in 30° and 90° flexion were 31.8°±1.2° and 39.2°±2.8°, respectively. The differences between the true and measured angles were not significant. The measured angle increased by more than 10° after cutting of the popliteus tendon in both 30° and 90° flexion. CONCLUSIONS: Using a smartphone application could be a good method of measuring knee rotation.