Cargando…
Angular dose dependency of MatriXX TM and its calibration
One of the applications of MatriXX (IBA Dosimetry) is experimental verification of dose for IMRT, VMAT, and tomotherapy. For cumulative plan verification, dose is delivered for all the treatment gantry angles to a stationary detector. Experimental calibration of MatriXX detector recommended by the m...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5719776/ https://www.ncbi.nlm.nih.gov/pubmed/20160692 http://dx.doi.org/10.1120/jacmp.v11i1.3057 |
_version_ | 1783284555300470784 |
---|---|
author | Wolfsberger, Luciant D. Wagar, Matthew Nitsch, Paige Bhagwat, Mandar S. Zygmanski, Piotr |
author_facet | Wolfsberger, Luciant D. Wagar, Matthew Nitsch, Paige Bhagwat, Mandar S. Zygmanski, Piotr |
author_sort | Wolfsberger, Luciant D. |
collection | PubMed |
description | One of the applications of MatriXX (IBA Dosimetry) is experimental verification of dose for IMRT, VMAT, and tomotherapy. For cumulative plan verification, dose is delivered for all the treatment gantry angles to a stationary detector. Experimental calibration of MatriXX detector recommended by the manufacturer involves only AP calibration fields and does not address angular dependency of MatriXX. Angular dependency may introduce dose bias in cumulative plan verification if not corrected. For this reason, we characterized angular dependency of MatriXX and developed a method for its calibration. We found relatively large discrepancies in responses to posterior vs. anterior fields for four MatriXX (Evolution series) detectors (up to 11%), and relatively large variability of responses as a function of gantry angle in the gantry angle ranges of 91°–110° and 269°–260°. With our calibration method, the bias due to angular dependency is effectively removed in experimental verification of IMRT and VMAT plans. PACS number: 87.56Fc |
format | Online Article Text |
id | pubmed-5719776 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57197762018-04-02 Angular dose dependency of MatriXX TM and its calibration Wolfsberger, Luciant D. Wagar, Matthew Nitsch, Paige Bhagwat, Mandar S. Zygmanski, Piotr J Appl Clin Med Phys Radiation Measurements One of the applications of MatriXX (IBA Dosimetry) is experimental verification of dose for IMRT, VMAT, and tomotherapy. For cumulative plan verification, dose is delivered for all the treatment gantry angles to a stationary detector. Experimental calibration of MatriXX detector recommended by the manufacturer involves only AP calibration fields and does not address angular dependency of MatriXX. Angular dependency may introduce dose bias in cumulative plan verification if not corrected. For this reason, we characterized angular dependency of MatriXX and developed a method for its calibration. We found relatively large discrepancies in responses to posterior vs. anterior fields for four MatriXX (Evolution series) detectors (up to 11%), and relatively large variability of responses as a function of gantry angle in the gantry angle ranges of 91°–110° and 269°–260°. With our calibration method, the bias due to angular dependency is effectively removed in experimental verification of IMRT and VMAT plans. PACS number: 87.56Fc John Wiley and Sons Inc. 2010-01-28 /pmc/articles/PMC5719776/ /pubmed/20160692 http://dx.doi.org/10.1120/jacmp.v11i1.3057 Text en © 2010 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Measurements Wolfsberger, Luciant D. Wagar, Matthew Nitsch, Paige Bhagwat, Mandar S. Zygmanski, Piotr Angular dose dependency of MatriXX TM and its calibration |
title | Angular dose dependency of MatriXX TM and its calibration |
title_full | Angular dose dependency of MatriXX TM and its calibration |
title_fullStr | Angular dose dependency of MatriXX TM and its calibration |
title_full_unstemmed | Angular dose dependency of MatriXX TM and its calibration |
title_short | Angular dose dependency of MatriXX TM and its calibration |
title_sort | angular dose dependency of matrixx tm and its calibration |
topic | Radiation Measurements |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5719776/ https://www.ncbi.nlm.nih.gov/pubmed/20160692 http://dx.doi.org/10.1120/jacmp.v11i1.3057 |
work_keys_str_mv | AT wolfsbergerluciantd angulardosedependencyofmatrixxtmanditscalibration AT wagarmatthew angulardosedependencyofmatrixxtmanditscalibration AT nitschpaige angulardosedependencyofmatrixxtmanditscalibration AT bhagwatmandars angulardosedependencyofmatrixxtmanditscalibration AT zygmanskipiotr angulardosedependencyofmatrixxtmanditscalibration |