High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction

To report a study of the use of GAFCHROMIC EBT radiochromic film (RCF) digitized with a commercially available flatbed document scanner for accurate and reliable all‐purpose two‐dimensional (2D) absolute dosimetry within a clinical environment. We used a simplified methodology that yields high‐preci...

Descripción completa

Detalles Bibliográficos
Autores principales: Chung, Heeteak, Lynch, Bart, Samant, Sanjiv
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5719957/
https://www.ncbi.nlm.nih.gov/pubmed/20592696
http://dx.doi.org/10.1120/jacmp.v11i2.3112
_version_ 1783284588009750528
author Chung, Heeteak
Lynch, Bart
Samant, Sanjiv
author_facet Chung, Heeteak
Lynch, Bart
Samant, Sanjiv
author_sort Chung, Heeteak
collection PubMed
description To report a study of the use of GAFCHROMIC EBT radiochromic film (RCF) digitized with a commercially available flatbed document scanner for accurate and reliable all‐purpose two‐dimensional (2D) absolute dosimetry within a clinical environment. We used a simplified methodology that yields high‐precision dosimetry measurements without significant postirradiation correction. The Epson Expression 1680 Professional scanner and the Epson Expression 10000XL scanner were used to digitize the films. Both scanners were retrofitted with light‐diffusing glass to minimize the effects of Newton rings. Known doses were delivered to calibration films. Flat and wedge fields were irradiated with variable depth of solid water and 5 cm back scatter solid water. No particular scanner nonuniformity effect corrections or significant post‐scan image processing were carried out. The profiles were compared with CC04 ionization chamber profiles. The depth dose distribution was measured at a source‐to‐surface distance (SSD) of 100 cm with a field size of [Formula: see text]. Additionally, 22 IMRT fields were measured and evaluated using gamma index analysis. The overall accuracy of RCF with respect to CC04 was found to be 2%–4%. The overall accuracy of RCF was determined using the absolute mean of difference for all flat and wedge field profiles. For clinical IMRT fields, both scanners showed an overall gamma index passing rate greater than 90%. This work demonstrated that EBT films, in conjunction with a commercially available flatbed scanner, can be used as an accurate and precise absolute dosimeter. Both scanners showed that no significant scanner nonuniformity correction is necessary for accurate absolute dosimetry using the EBT films for field sizes smaller than or equal to [Formula: see text]. PACS number: 87.53.Bn
format Online
Article
Text
id pubmed-5719957
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57199572018-04-02 High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction Chung, Heeteak Lynch, Bart Samant, Sanjiv J Appl Clin Med Phys Radiation Oncology Physics To report a study of the use of GAFCHROMIC EBT radiochromic film (RCF) digitized with a commercially available flatbed document scanner for accurate and reliable all‐purpose two‐dimensional (2D) absolute dosimetry within a clinical environment. We used a simplified methodology that yields high‐precision dosimetry measurements without significant postirradiation correction. The Epson Expression 1680 Professional scanner and the Epson Expression 10000XL scanner were used to digitize the films. Both scanners were retrofitted with light‐diffusing glass to minimize the effects of Newton rings. Known doses were delivered to calibration films. Flat and wedge fields were irradiated with variable depth of solid water and 5 cm back scatter solid water. No particular scanner nonuniformity effect corrections or significant post‐scan image processing were carried out. The profiles were compared with CC04 ionization chamber profiles. The depth dose distribution was measured at a source‐to‐surface distance (SSD) of 100 cm with a field size of [Formula: see text]. Additionally, 22 IMRT fields were measured and evaluated using gamma index analysis. The overall accuracy of RCF with respect to CC04 was found to be 2%–4%. The overall accuracy of RCF was determined using the absolute mean of difference for all flat and wedge field profiles. For clinical IMRT fields, both scanners showed an overall gamma index passing rate greater than 90%. This work demonstrated that EBT films, in conjunction with a commercially available flatbed scanner, can be used as an accurate and precise absolute dosimeter. Both scanners showed that no significant scanner nonuniformity correction is necessary for accurate absolute dosimetry using the EBT films for field sizes smaller than or equal to [Formula: see text]. PACS number: 87.53.Bn John Wiley and Sons Inc. 2010-04-17 /pmc/articles/PMC5719957/ /pubmed/20592696 http://dx.doi.org/10.1120/jacmp.v11i2.3112 Text en © 2010 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Chung, Heeteak
Lynch, Bart
Samant, Sanjiv
High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
title High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
title_full High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
title_fullStr High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
title_full_unstemmed High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
title_short High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
title_sort high‐precision gafchromic ebt film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5719957/
https://www.ncbi.nlm.nih.gov/pubmed/20592696
http://dx.doi.org/10.1120/jacmp.v11i2.3112
work_keys_str_mv AT chungheeteak highprecisiongafchromicebtfilmbasedabsoluteclinicaldosimetryusingastandardflatbedscannerwithouttheuseofascannernonuniformitycorrection
AT lynchbart highprecisiongafchromicebtfilmbasedabsoluteclinicaldosimetryusingastandardflatbedscannerwithouttheuseofascannernonuniformitycorrection
AT samantsanjiv highprecisiongafchromicebtfilmbasedabsoluteclinicaldosimetryusingastandardflatbedscannerwithouttheuseofascannernonuniformitycorrection