A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator

A quality assurance (QA) procedure was developed to evaluate the congruence between the cone‐beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be p...

Descripción completa

Detalles Bibliográficos
Autores principales: Du, Weiliang, Yang, James N., Chang, Eric L., Luo, Dershan, McAleer, Mary Frances, Shiu, Almon, Martel, Mary K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720414/
https://www.ncbi.nlm.nih.gov/pubmed/21081890
http://dx.doi.org/10.1120/jacmp.v11i4.3297
_version_ 1783284643878928384
author Du, Weiliang
Yang, James N.
Chang, Eric L.
Luo, Dershan
McAleer, Mary Frances
Shiu, Almon
Martel, Mary K.
author_facet Du, Weiliang
Yang, James N.
Chang, Eric L.
Luo, Dershan
McAleer, Mary Frances
Shiu, Almon
Martel, Mary K.
author_sort Du, Weiliang
collection PubMed
description A quality assurance (QA) procedure was developed to evaluate the congruence between the cone‐beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be placed exactly at the radiation isocenter through precalibrated room lasers or light field crosshairs. The only requirement was that the BB phantom be in a stationary position near the radiation isocenter during the image acquisition process. The radiation isocenter was determined with respect to the center of the BB using a Winston‐Lutz test. The CBCT image center was found to have excellent short‐term positional reproducibility (i.e., less than 0.1 mm of wobble in each of the x (lateral), y (vertical), and z (longitudinal) directions) in 10 consecutive acquisitions. Measured over a seven‐month period, the CBCT image center deviated from the radiation isocenter by [Formula: see text] , and [Formula: see text]. The z displacement of the 3D CBCT image center was highly correlated [Formula: see text] with that of the 2D kV portal image center. The correlation coefficients in the x and y directions were poor ([Formula: see text] and ‐0.35, respectively). Systematic discrepancies were found between the CBCT image center and the 2D MV, kV portal image centers. For the linear accelerator studied, we detected a 0.8 mm discrepancy between the CBCT image center and the MV EPID image center in the anterior‐posterior direction. This discrepancy was demonstrated in a clinical case study where the patient was positioned with CBCT followed by MV portal verification. The results from the new QA procedure are useful for guiding high‐precision patient positioning in stereotactic body radiation therapy. PACS number: 87.55.Qr
format Online
Article
Text
id pubmed-5720414
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57204142018-04-02 A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator Du, Weiliang Yang, James N. Chang, Eric L. Luo, Dershan McAleer, Mary Frances Shiu, Almon Martel, Mary K. J Appl Clin Med Phys Radiation Oncology Physics A quality assurance (QA) procedure was developed to evaluate the congruence between the cone‐beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be placed exactly at the radiation isocenter through precalibrated room lasers or light field crosshairs. The only requirement was that the BB phantom be in a stationary position near the radiation isocenter during the image acquisition process. The radiation isocenter was determined with respect to the center of the BB using a Winston‐Lutz test. The CBCT image center was found to have excellent short‐term positional reproducibility (i.e., less than 0.1 mm of wobble in each of the x (lateral), y (vertical), and z (longitudinal) directions) in 10 consecutive acquisitions. Measured over a seven‐month period, the CBCT image center deviated from the radiation isocenter by [Formula: see text] , and [Formula: see text]. The z displacement of the 3D CBCT image center was highly correlated [Formula: see text] with that of the 2D kV portal image center. The correlation coefficients in the x and y directions were poor ([Formula: see text] and ‐0.35, respectively). Systematic discrepancies were found between the CBCT image center and the 2D MV, kV portal image centers. For the linear accelerator studied, we detected a 0.8 mm discrepancy between the CBCT image center and the MV EPID image center in the anterior‐posterior direction. This discrepancy was demonstrated in a clinical case study where the patient was positioned with CBCT followed by MV portal verification. The results from the new QA procedure are useful for guiding high‐precision patient positioning in stereotactic body radiation therapy. PACS number: 87.55.Qr John Wiley and Sons Inc. 2010-07-02 /pmc/articles/PMC5720414/ /pubmed/21081890 http://dx.doi.org/10.1120/jacmp.v11i4.3297 Text en © 2010 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Du, Weiliang
Yang, James N.
Chang, Eric L.
Luo, Dershan
McAleer, Mary Frances
Shiu, Almon
Martel, Mary K.
A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator
title A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator
title_full A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator
title_fullStr A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator
title_full_unstemmed A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator
title_short A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator
title_sort quality assurance procedure to evaluate cone‐beam ct image center congruence with the radiation isocenter of a linear accelerator
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720414/
https://www.ncbi.nlm.nih.gov/pubmed/21081890
http://dx.doi.org/10.1120/jacmp.v11i4.3297
work_keys_str_mv AT duweiliang aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT yangjamesn aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT changericl aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT luodershan aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT mcaleermaryfrances aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT shiualmon aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT martelmaryk aqualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT duweiliang qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT yangjamesn qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT changericl qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT luodershan qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT mcaleermaryfrances qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT shiualmon qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator
AT martelmaryk qualityassuranceproceduretoevaluateconebeamctimagecentercongruencewiththeradiationisocenterofalinearaccelerator