A quality assurance (QA) procedure was developed to evaluate the congruence between the cone‐beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be p...
Autores principales: | Du, Weiliang, Yang, James N., Chang, Eric L., Luo, Dershan, McAleer, Mary Frances, Shiu, Almon, Martel, Mary K. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720414/ https://www.ncbi.nlm.nih.gov/pubmed/21081890 http://dx.doi.org/10.1120/jacmp.v11i4.3297 |
Ejemplares similares
-
Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence
por: Kang, Hyejoo, et al.
Publicado: (2019) -
The spatial accuracy of two frameless, linear accelerator‐based systems for single‐isocenter, multitarget cranial radiosurgery
por: Ezzell, Gary A.
Publicado: (2017) -
Improved accuracy for noncoplanar radiotherapy: an EPID‐based method for submillimeter alignment of linear accelerator table rotation with MV isocenter
por: Nyflot, Matthew J., et al.
Publicado: (2014) -
Automatic detection of graticule isocenter and scale from kV and MV images
por: Fang, Raymond, et al.
Publicado: (2019) -
Dosimetric comparison of multiple vs single isocenter technique for linear accelerator‐based stereotactic radiosurgery: The Importance of the six degree couch
por: Amaya, Dania, et al.
Publicado: (2021)