Cargando…
Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation
We design a practical procedure for measuring translational and rotational misalignment of graticule with collimator axis of rotation and collimator jaws, respectively. The procedure's quantitative results are accurate to less than 0.2 mm (at isocenter) and do not assume alignment of radiation...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720415/ https://www.ncbi.nlm.nih.gov/pubmed/21081892 http://dx.doi.org/10.1120/jacmp.v11i4.3318 |
_version_ | 1783284644112760832 |
---|---|
author | Halabi, Tarek Faddegon, Bruce |
author_facet | Halabi, Tarek Faddegon, Bruce |
author_sort | Halabi, Tarek |
collection | PubMed |
description | We design a practical procedure for measuring translational and rotational misalignment of graticule with collimator axis of rotation and collimator jaws, respectively. The procedure's quantitative results are accurate to less than 0.2 mm (at isocenter) and do not assume alignment of radiation focal spot with collimator axis of rotation. When provided with these quantitative results, the manufacturer can custom‐adjust graticules to the purchaser's collimator head. PACS number: 87.56.Fc |
format | Online Article Text |
id | pubmed-5720415 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57204152018-04-02 Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation Halabi, Tarek Faddegon, Bruce J Appl Clin Med Phys Technical Notes We design a practical procedure for measuring translational and rotational misalignment of graticule with collimator axis of rotation and collimator jaws, respectively. The procedure's quantitative results are accurate to less than 0.2 mm (at isocenter) and do not assume alignment of radiation focal spot with collimator axis of rotation. When provided with these quantitative results, the manufacturer can custom‐adjust graticules to the purchaser's collimator head. PACS number: 87.56.Fc John Wiley and Sons Inc. 2010-08-15 /pmc/articles/PMC5720415/ /pubmed/21081892 http://dx.doi.org/10.1120/jacmp.v11i4.3318 Text en © 2010 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Notes Halabi, Tarek Faddegon, Bruce Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
title | Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
title_full | Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
title_fullStr | Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
title_full_unstemmed | Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
title_short | Practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
title_sort | practical quantitative measurement of graticule misalignment relative to collimator axis of rotation |
topic | Technical Notes |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720415/ https://www.ncbi.nlm.nih.gov/pubmed/21081892 http://dx.doi.org/10.1120/jacmp.v11i4.3318 |
work_keys_str_mv | AT halabitarek practicalquantitativemeasurementofgraticulemisalignmentrelativetocollimatoraxisofrotation AT faddegonbruce practicalquantitativemeasurementofgraticulemisalignmentrelativetocollimatoraxisofrotation |