A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number

The objective of this study was to evaluate the effect of kilovoltage cone‐beam computed tomography (CBCT) on registration accuracy and image qualities with a reduced number of planar projections used in volumetric imaging reconstruction. The ultimate goal is to evaluate the possibility of reducing...

Descripción completa

Detalles Bibliográficos
Autores principales: Lu, Bo, Lu, Haibin, Palta, Jatinder
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720437/
https://www.ncbi.nlm.nih.gov/pubmed/20717096
http://dx.doi.org/10.1120/jacmp.v11i3.3274
_version_ 1783284648946696192
author Lu, Bo
Lu, Haibin
Palta, Jatinder
author_facet Lu, Bo
Lu, Haibin
Palta, Jatinder
author_sort Lu, Bo
collection PubMed
description The objective of this study was to evaluate the effect of kilovoltage cone‐beam computed tomography (CBCT) on registration accuracy and image qualities with a reduced number of planar projections used in volumetric imaging reconstruction. The ultimate goal is to evaluate the possibility of reducing the patient dose while maintaining registration accuracy under different projection‐number schemes for various clinical sites. An Elekta Synergy Linear accelerator with an onboard CBCT system was used in this study. The quality of the Elekta XVI cone‐beam three‐dimensional volumetric images reconstructed with a decreasing number of projections was quantitatively evaluated by a Catphan phantom. Subsequently, we tested the registration accuracy of imaging data sets on three rigid anthropomorphic phantoms and three real patient sites under the reduced projection‐number (as low as 1/6th) reconstruction of CBCT data with different rectilinear shifts and rotations. CBCT scan results of the Catphan phantom indicated the CBCT images got noisier when the number of projections was reduced, but their spatial resolution and uniformity were hardly affected. The maximum registration errors under the small amount transformation of the reference CT images were found to be within 0.7 mm translation and 0.3° rotation. However, when the projection number was lower than one‐fourth of the full set with a large amount of transformation of reference CT images, the registration could easily be trapped into local minima solutions for a nonrigid anatomy. We concluded, by using projection‐number reduction strategy under conscientious care, imaging‐guided localization procedure could achieve a lower patient dose without losing the registration accuracy for various clinical sites and situations. A faster scanning time is the main advantage compared to the mA decrease‐based, dose‐reduction method. PACS numbers: 87.57.C‐, 87.57.cf, 87.57.cj, 87.57.cm, 87.57.cp, 87.57.N‐, 87.57.nf, 87.57.nj
format Online
Article
Text
id pubmed-5720437
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57204372018-04-02 A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number Lu, Bo Lu, Haibin Palta, Jatinder J Appl Clin Med Phys Medical Imaging The objective of this study was to evaluate the effect of kilovoltage cone‐beam computed tomography (CBCT) on registration accuracy and image qualities with a reduced number of planar projections used in volumetric imaging reconstruction. The ultimate goal is to evaluate the possibility of reducing the patient dose while maintaining registration accuracy under different projection‐number schemes for various clinical sites. An Elekta Synergy Linear accelerator with an onboard CBCT system was used in this study. The quality of the Elekta XVI cone‐beam three‐dimensional volumetric images reconstructed with a decreasing number of projections was quantitatively evaluated by a Catphan phantom. Subsequently, we tested the registration accuracy of imaging data sets on three rigid anthropomorphic phantoms and three real patient sites under the reduced projection‐number (as low as 1/6th) reconstruction of CBCT data with different rectilinear shifts and rotations. CBCT scan results of the Catphan phantom indicated the CBCT images got noisier when the number of projections was reduced, but their spatial resolution and uniformity were hardly affected. The maximum registration errors under the small amount transformation of the reference CT images were found to be within 0.7 mm translation and 0.3° rotation. However, when the projection number was lower than one‐fourth of the full set with a large amount of transformation of reference CT images, the registration could easily be trapped into local minima solutions for a nonrigid anatomy. We concluded, by using projection‐number reduction strategy under conscientious care, imaging‐guided localization procedure could achieve a lower patient dose without losing the registration accuracy for various clinical sites and situations. A faster scanning time is the main advantage compared to the mA decrease‐based, dose‐reduction method. PACS numbers: 87.57.C‐, 87.57.cf, 87.57.cj, 87.57.cm, 87.57.cp, 87.57.N‐, 87.57.nf, 87.57.nj John Wiley and Sons Inc. 2010-05-12 /pmc/articles/PMC5720437/ /pubmed/20717096 http://dx.doi.org/10.1120/jacmp.v11i3.3274 Text en © 2010 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Lu, Bo
Lu, Haibin
Palta, Jatinder
A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number
title A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number
title_full A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number
title_fullStr A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number
title_full_unstemmed A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number
title_short A comprehensive study on decreasing the kilovoltage cone‐beam CT dose by reducing the projection number
title_sort comprehensive study on decreasing the kilovoltage cone‐beam ct dose by reducing the projection number
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720437/
https://www.ncbi.nlm.nih.gov/pubmed/20717096
http://dx.doi.org/10.1120/jacmp.v11i3.3274
work_keys_str_mv AT lubo acomprehensivestudyondecreasingthekilovoltageconebeamctdosebyreducingtheprojectionnumber
AT luhaibin acomprehensivestudyondecreasingthekilovoltageconebeamctdosebyreducingtheprojectionnumber
AT paltajatinder acomprehensivestudyondecreasingthekilovoltageconebeamctdosebyreducingtheprojectionnumber
AT lubo comprehensivestudyondecreasingthekilovoltageconebeamctdosebyreducingtheprojectionnumber
AT luhaibin comprehensivestudyondecreasingthekilovoltageconebeamctdosebyreducingtheprojectionnumber
AT paltajatinder comprehensivestudyondecreasingthekilovoltageconebeamctdosebyreducingtheprojectionnumber