Leakage reduction for the Siemens ModuLeaf
The ModuLeaf, an add‐on miniature multileaf collimator (MMLC) for the Siemens ONCOR linear accelerator, provides high resolution field shaping with a maximum interleaf leakage dose of 1.50% at 6 MV. However, beyond the maximum treatment field size, the distribution of leakage and scatter along the y...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720452/ https://www.ncbi.nlm.nih.gov/pubmed/19458591 http://dx.doi.org/10.1120/jacmp.v10i2.2894 |
Sumario: | The ModuLeaf, an add‐on miniature multileaf collimator (MMLC) for the Siemens ONCOR linear accelerator, provides high resolution field shaping with a maximum interleaf leakage dose of 1.50% at 6 MV. However, beyond the maximum treatment field size, the distribution of leakage and scatter along the y‐axis is different from that of the x‐axis, with maximum leakage values of 1.53% and 0.39%, respectively. Such differences cannot be modeled in the Pinnacle treatment planning system. Also, within the [Formula: see text] treatment region, leakage from the crack between closed leaf ends was 3.76%. To resolve these issues, gaps in the ModuLeaf frame were filled with lead sheets, and the Siemens MLC was operated in MLC mode (rather than bank mode). As a result, a rectangle of [Formula: see text] was formed with the MLC leaves closed behind the Y jaws, whose opening was 10.4 cm. This significantly reduced the difference between the leakage patterns in the x and y directions, with maximum leakage doses of 0.43% outside the treatment region and 1.67% near the crack between abutting ModuLeaf leaves. The modification also reduced the mean square error between Pinnacle profiles and measured profiles in the tail region. PACS number: 87.56.jk |
---|