Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes
Megavoltage cone‐beam CT (MVCBCT) is an imaging technology that provides a 3D representation of the patient in treatment position. Because it is a form of x‐ray tomography, MVCBCT images give information about the attenuation coefficients of the imaged tissues, and thus could be used for dose calcul...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720499/ https://www.ncbi.nlm.nih.gov/pubmed/19223832 http://dx.doi.org/10.1120/jacmp.v10i1.2852 |
_version_ | 1783284660748419072 |
---|---|
author | Aubry, Jean‐François Cheung, Joey Morin, Olivier Gottschalk, Alexander Beaulieu, Luc Pouliot, Jean |
author_facet | Aubry, Jean‐François Cheung, Joey Morin, Olivier Gottschalk, Alexander Beaulieu, Luc Pouliot, Jean |
author_sort | Aubry, Jean‐François |
collection | PubMed |
description | Megavoltage cone‐beam CT (MVCBCT) is an imaging technology that provides a 3D representation of the patient in treatment position. Because it is a form of x‐ray tomography, MVCBCT images give information about the attenuation coefficients of the imaged tissues, and thus could be used for dose calculation. However, the cupping and missing data artifacts seen on MVCBCT images can cause inaccuracies in dose calculations. To eliminate these inaccuracies, a correction method specific to pelvis imaging and based on phantom measurements has been devised. Pelvis‐shaped water phantoms of three different sizes were designed and imaged with MVCBCT. Three sets of correction factors were created from the artifacts observed in these MVCBCT images by dividing the measured CT number by the predefined CT number for water. Linear interpolation is performed between the sets of correction factors to take into account the varying size of different patients. To compensate for the missing anatomy due to the limited field of view of the MVCBCT system, the MVCBCT image is complemented with the kilovoltage CT (kVCT) image acquired for treatment planning. When the correction method is applied to an anthropomorphic pelvis phantom, the standard deviation between dose calculations performed with kVCT and MVCBCT images is 0.6%, with 98% of the dose points agreeing within [Formula: see text]. With uncorrected MVCBCT images this percentage falls to 75%. An example of dose calculation performed with a corrected clinical MVCBCT image of a prostate cancer patient shows that changes in anatomy of normal tissues result in variation of the dose distribution received by these tissues. This correction method enables MVCBCT images to be used for the verification of the daily dose distribution for patients treated in the pelvis region. PACS numbers: 87.57.Q‐ Computed tomography, 87.57.cp Artifacts and distortion |
format | Online Article Text |
id | pubmed-5720499 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2009 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57204992018-04-02 Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes Aubry, Jean‐François Cheung, Joey Morin, Olivier Gottschalk, Alexander Beaulieu, Luc Pouliot, Jean J Appl Clin Med Phys Radiation Oncology Physics Megavoltage cone‐beam CT (MVCBCT) is an imaging technology that provides a 3D representation of the patient in treatment position. Because it is a form of x‐ray tomography, MVCBCT images give information about the attenuation coefficients of the imaged tissues, and thus could be used for dose calculation. However, the cupping and missing data artifacts seen on MVCBCT images can cause inaccuracies in dose calculations. To eliminate these inaccuracies, a correction method specific to pelvis imaging and based on phantom measurements has been devised. Pelvis‐shaped water phantoms of three different sizes were designed and imaged with MVCBCT. Three sets of correction factors were created from the artifacts observed in these MVCBCT images by dividing the measured CT number by the predefined CT number for water. Linear interpolation is performed between the sets of correction factors to take into account the varying size of different patients. To compensate for the missing anatomy due to the limited field of view of the MVCBCT system, the MVCBCT image is complemented with the kilovoltage CT (kVCT) image acquired for treatment planning. When the correction method is applied to an anthropomorphic pelvis phantom, the standard deviation between dose calculations performed with kVCT and MVCBCT images is 0.6%, with 98% of the dose points agreeing within [Formula: see text]. With uncorrected MVCBCT images this percentage falls to 75%. An example of dose calculation performed with a corrected clinical MVCBCT image of a prostate cancer patient shows that changes in anatomy of normal tissues result in variation of the dose distribution received by these tissues. This correction method enables MVCBCT images to be used for the verification of the daily dose distribution for patients treated in the pelvis region. PACS numbers: 87.57.Q‐ Computed tomography, 87.57.cp Artifacts and distortion John Wiley and Sons Inc. 2009-01-27 /pmc/articles/PMC5720499/ /pubmed/19223832 http://dx.doi.org/10.1120/jacmp.v10i1.2852 Text en © 2009 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Aubry, Jean‐François Cheung, Joey Morin, Olivier Gottschalk, Alexander Beaulieu, Luc Pouliot, Jean Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes |
title | Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes |
title_full | Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes |
title_fullStr | Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes |
title_full_unstemmed | Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes |
title_short | Correction of megavoltage cone‐beam CT images of the pelvic region based on phantom measurements for dose calculation purposes |
title_sort | correction of megavoltage cone‐beam ct images of the pelvic region based on phantom measurements for dose calculation purposes |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720499/ https://www.ncbi.nlm.nih.gov/pubmed/19223832 http://dx.doi.org/10.1120/jacmp.v10i1.2852 |
work_keys_str_mv | AT aubryjeanfrancois correctionofmegavoltageconebeamctimagesofthepelvicregionbasedonphantommeasurementsfordosecalculationpurposes AT cheungjoey correctionofmegavoltageconebeamctimagesofthepelvicregionbasedonphantommeasurementsfordosecalculationpurposes AT morinolivier correctionofmegavoltageconebeamctimagesofthepelvicregionbasedonphantommeasurementsfordosecalculationpurposes AT gottschalkalexander correctionofmegavoltageconebeamctimagesofthepelvicregionbasedonphantommeasurementsfordosecalculationpurposes AT beaulieuluc correctionofmegavoltageconebeamctimagesofthepelvicregionbasedonphantommeasurementsfordosecalculationpurposes AT pouliotjean correctionofmegavoltageconebeamctimagesofthepelvicregionbasedonphantommeasurementsfordosecalculationpurposes |