Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers

To investigate the effects of sagging shifts of three on‐board kV imaging systems (OBI) on the isocenter positioning accuracy and image quality of cone‐beam CT (CBCT). A cubical phantom having a metal marker in the center that can be aligned with the radiation isocenter was used to measure sagging s...

Descripción completa

Detalles Bibliográficos
Autores principales: Ali, Imad, Ahmad, Salahuddin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720558/
https://www.ncbi.nlm.nih.gov/pubmed/19692976
http://dx.doi.org/10.1120/jacmp.v10i3.2930
_version_ 1783284674752151552
author Ali, Imad
Ahmad, Salahuddin
author_facet Ali, Imad
Ahmad, Salahuddin
author_sort Ali, Imad
collection PubMed
description To investigate the effects of sagging shifts of three on‐board kV imaging systems (OBI) on the isocenter positioning accuracy and image quality of cone‐beam CT (CBCT). A cubical phantom having a metal marker in the center that can be aligned with the radiation isocenter was used to measure sagging shifts and their variation with gantry angle on three Varian linacs with kV on‐board imaging systems. A marker‐tracking algorithm was applied to detect the shadow of the metal marker and localize its center in the two‐dimensional cone‐beam radiographic projections. This tracking algorithm is based on finding the position of maximum cross‐correlation between a region‐of‐interest from a template image (including the metal marker) and the projections containing the shadow of the metal marker. Sagging shifts were corrected by mapping the center of the metal marker to a reference position for all projections acquired over a full gantry rotation (0–360°). The sag‐corrected radiographic projections were then used to reconstruct CBCT using Feldkamp back‐projection. A standard quality assurance phantom was used to evaluate the image quality of CBCT before and after sagging correction. Sagging affects both the positioning accuracy of the OBI isocenter and the CBCT image quality. For example, on one linac, the position of the marker on the cone‐beam radiographic projections depends on the angular view and has maximal shifts of about 2 mm along the imager x‐direction (patient's cross‐plane). Sagging produces systematic shifts of the OBI isocenter as large as 1 mm posterior and 1 mm left in patient coordinates relative to the radiation isocenter. Further, it causes spatial distortion and blurring in CBCT image reconstructed from radiographic projections that are not corrected for OBI sagging. CBCT numbers vary by about 1% in full‐fan scans and up to 3.5% in half‐fan scans because of sagging. In order to achieve better localization accuracy in image‐guided radiation therapy, sagging shifts of the kV OBI need to be corrected. In addition, correction of sagging improves image and provides better visualization of internal structures. Frequent quality assurance is required to monitor and maintain standards of variations in the mechanical accuracy of isocenter and image quality of CBCT because of sagging shifts. PACS number: 87.57.Q, 87.57.cp
format Online
Article
Text
id pubmed-5720558
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57205582018-04-02 Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers Ali, Imad Ahmad, Salahuddin J Appl Clin Med Phys Medical Imaging To investigate the effects of sagging shifts of three on‐board kV imaging systems (OBI) on the isocenter positioning accuracy and image quality of cone‐beam CT (CBCT). A cubical phantom having a metal marker in the center that can be aligned with the radiation isocenter was used to measure sagging shifts and their variation with gantry angle on three Varian linacs with kV on‐board imaging systems. A marker‐tracking algorithm was applied to detect the shadow of the metal marker and localize its center in the two‐dimensional cone‐beam radiographic projections. This tracking algorithm is based on finding the position of maximum cross‐correlation between a region‐of‐interest from a template image (including the metal marker) and the projections containing the shadow of the metal marker. Sagging shifts were corrected by mapping the center of the metal marker to a reference position for all projections acquired over a full gantry rotation (0–360°). The sag‐corrected radiographic projections were then used to reconstruct CBCT using Feldkamp back‐projection. A standard quality assurance phantom was used to evaluate the image quality of CBCT before and after sagging correction. Sagging affects both the positioning accuracy of the OBI isocenter and the CBCT image quality. For example, on one linac, the position of the marker on the cone‐beam radiographic projections depends on the angular view and has maximal shifts of about 2 mm along the imager x‐direction (patient's cross‐plane). Sagging produces systematic shifts of the OBI isocenter as large as 1 mm posterior and 1 mm left in patient coordinates relative to the radiation isocenter. Further, it causes spatial distortion and blurring in CBCT image reconstructed from radiographic projections that are not corrected for OBI sagging. CBCT numbers vary by about 1% in full‐fan scans and up to 3.5% in half‐fan scans because of sagging. In order to achieve better localization accuracy in image‐guided radiation therapy, sagging shifts of the kV OBI need to be corrected. In addition, correction of sagging improves image and provides better visualization of internal structures. Frequent quality assurance is required to monitor and maintain standards of variations in the mechanical accuracy of isocenter and image quality of CBCT because of sagging shifts. PACS number: 87.57.Q, 87.57.cp John Wiley and Sons Inc. 2009-07-17 /pmc/articles/PMC5720558/ /pubmed/19692976 http://dx.doi.org/10.1120/jacmp.v10i3.2930 Text en © 2009 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Ali, Imad
Ahmad, Salahuddin
Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers
title Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers
title_full Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers
title_fullStr Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers
title_full_unstemmed Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers
title_short Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam CT from kV on‐board imagers
title_sort evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone‐beam ct from kv on‐board imagers
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720558/
https://www.ncbi.nlm.nih.gov/pubmed/19692976
http://dx.doi.org/10.1120/jacmp.v10i3.2930
work_keys_str_mv AT aliimad evaluationoftheeffectsofsaggingshiftsonisocenteraccuracyandimagequalityofconebeamctfromkvonboardimagers
AT ahmadsalahuddin evaluationoftheeffectsofsaggingshiftsonisocenteraccuracyandimagequalityofconebeamctfromkvonboardimagers