A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on‐board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pr...

Descripción completa

Detalles Bibliográficos
Autores principales: Guan, Huaiqun, Hammoud, Rabih, Yin, Fang‐Fang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720565/
https://www.ncbi.nlm.nih.gov/pubmed/19918223
http://dx.doi.org/10.1120/jacmp.v10i4.2954
_version_ 1783284676453990400
author Guan, Huaiqun
Hammoud, Rabih
Yin, Fang‐Fang
author_facet Guan, Huaiqun
Hammoud, Rabih
Yin, Fang‐Fang
author_sort Guan, Huaiqun
collection PubMed
description A positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on‐board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pre‐specified accuracy); (2) that the positioning difference detected by 2D/2D (kV/MV) and 3D/3D(planCT/CBCT) matching can be reliably transferred to couch motion. A cube phantom with a 2 mm metal ball (bb) at the center was used. The bb was used to define the isocenter. Two additional bbs were placed on two phantom surfaces in order to define a spatial location of 1.5 cm anterior, 1.5 cm inferior, and 1.5 cm right from the isocenter. An axial scan of the phantom was acquired from a multislice CT simulator. The phantom was set at the linac's isocenter (lasers); either AP MV/R Lat kV images or CBCT images were taken for 2D/2D or 3D/3D matching, respectively. For 2D/2D, the accuracy of each device's isocenter was obtained by checking the distance between the central bb and the digital graticule. Then the central bb in orthogonal DRRs was manually moved to overlay to the off‐axis bbs in kV/MV images. For 3D/3D, CBCT was first matched to planCT to check the isocenter difference between the two CTs. Manual shifts were then made by moving CBCT such that the point defined by the two off‐axis bbs overlay to the central bb in planCT. (PlanCT can not be moved in the current version of OBI1.4.) The manual shifts were then applied to remotely move the couch. The room laser was used to check the accuracy of the couch movement. For Trilogy (or Ix‐21) linacs, the coincidence of imager and linac's isocenter was better than 1 mm (or 1.5 mm). The couch shift accuracy was better than 2 mm. PACS number: 87.55.Qr, 87.57.Q‐
format Online
Article
Text
id pubmed-5720565
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57205652018-04-02 A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Guan, Huaiqun Hammoud, Rabih Yin, Fang‐Fang J Appl Clin Med Phys Technical Notes A positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on‐board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pre‐specified accuracy); (2) that the positioning difference detected by 2D/2D (kV/MV) and 3D/3D(planCT/CBCT) matching can be reliably transferred to couch motion. A cube phantom with a 2 mm metal ball (bb) at the center was used. The bb was used to define the isocenter. Two additional bbs were placed on two phantom surfaces in order to define a spatial location of 1.5 cm anterior, 1.5 cm inferior, and 1.5 cm right from the isocenter. An axial scan of the phantom was acquired from a multislice CT simulator. The phantom was set at the linac's isocenter (lasers); either AP MV/R Lat kV images or CBCT images were taken for 2D/2D or 3D/3D matching, respectively. For 2D/2D, the accuracy of each device's isocenter was obtained by checking the distance between the central bb and the digital graticule. Then the central bb in orthogonal DRRs was manually moved to overlay to the off‐axis bbs in kV/MV images. For 3D/3D, CBCT was first matched to planCT to check the isocenter difference between the two CTs. Manual shifts were then made by moving CBCT such that the point defined by the two off‐axis bbs overlay to the central bb in planCT. (PlanCT can not be moved in the current version of OBI1.4.) The manual shifts were then applied to remotely move the couch. The room laser was used to check the accuracy of the couch movement. For Trilogy (or Ix‐21) linacs, the coincidence of imager and linac's isocenter was better than 1 mm (or 1.5 mm). The couch shift accuracy was better than 2 mm. PACS number: 87.55.Qr, 87.57.Q‐ John Wiley and Sons Inc. 2009-10-06 /pmc/articles/PMC5720565/ /pubmed/19918223 http://dx.doi.org/10.1120/jacmp.v10i4.2954 Text en © 2009 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Guan, Huaiqun
Hammoud, Rabih
Yin, Fang‐Fang
A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
title A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
title_full A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
title_fullStr A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
title_full_unstemmed A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
title_short A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
title_sort positioning qa procedure for 2d/2d (kv/mv) and 3d/3d (ct/cbct) image matching for radiotherapy patient setup
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720565/
https://www.ncbi.nlm.nih.gov/pubmed/19918223
http://dx.doi.org/10.1120/jacmp.v10i4.2954
work_keys_str_mv AT guanhuaiqun apositioningqaprocedurefor2d2dkvmvand3d3dctcbctimagematchingforradiotherapypatientsetup
AT hammoudrabih apositioningqaprocedurefor2d2dkvmvand3d3dctcbctimagematchingforradiotherapypatientsetup
AT yinfangfang apositioningqaprocedurefor2d2dkvmvand3d3dctcbctimagematchingforradiotherapypatientsetup
AT guanhuaiqun positioningqaprocedurefor2d2dkvmvand3d3dctcbctimagematchingforradiotherapypatientsetup
AT hammoudrabih positioningqaprocedurefor2d2dkvmvand3d3dctcbctimagematchingforradiotherapypatientsetup
AT yinfangfang positioningqaprocedurefor2d2dkvmvand3d3dctcbctimagematchingforradiotherapypatientsetup