Evaluation of a biplanar diode array dosimeter for quality assurance of step‐and‐shoot IMRT

In this paper, we describe and characterize a novel biplanar diode array, and demonstrate its applicability to dosimetric QA of step‐and‐shoot IMRT. It is the first commercially available device of its kind specifically designed for performing measurements at varying gantry angles. The detector cons...

Descripción completa

Detalles Bibliográficos
Autores principales: Feygelman, Vladimir, Forster, Kenneth, Opp, Daniel, Nilsson, Görgen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720580/
https://www.ncbi.nlm.nih.gov/pubmed/19918238
http://dx.doi.org/10.1120/jacmp.v10i4.3080
Descripción
Sumario:In this paper, we describe and characterize a novel biplanar diode array, and demonstrate its applicability to dosimetric QA of step‐and‐shoot IMRT. It is the first commercially available device of its kind specifically designed for performing measurements at varying gantry angles. The detector consists of a cylindrical PMMA phantom with two orthogonal detector boards. There are a total of 1069 p‐type 1 mm wide diode detectors covering the measurement area of [Formula: see text] in each of the measurement planes. The orthogonal detector arrays ensure that the dose modulation information is not lost regardless of the beam incidence angle. For absolute calibration, the dose to the reference detector is calculated at the appropriate SSD and radiological depth by the treatment planning system and is scaled by the measured accelerator output. The directly measured rotational response on the central axis shows the maximum variation of approximately [Formula: see text] in the narrow [Formula: see text] angular intervals centered on the detector boards. This variation is reduced to less than [Formula: see text] outside of the four similarly centered [Formula: see text] angular intervals. For all detectors, the difference between the measured and the calculated dose for a plan with 12 equally spaced beams is [Formula: see text]. Of eleven IMRT plans, ten passed the γ (3%,3 mm) criterion at or above 95%, while one passed at 92%. The biplanar diode array is a useful tool for IMRT QA, allowing for essentially instantaneous online analysis of absolute dose errors in 3D. PACS number: 87.55Qr