Cargando…
Effect of output variation with dose rate on the Virtual Wedge factor
The Siemens Virtual Wedge factor (VWF: Siemens Medical Solutions, Malvern, PA) may drift significantly because of an increase in output as the dose rate declines. This variation in output is caused by sample and hold boards in the dosimetry circuit that become defective because of radiation damage....
Autores principales: | Zhu, X. Ronald, Gillin, Michael T. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5721535/ https://www.ncbi.nlm.nih.gov/pubmed/18449167 http://dx.doi.org/10.1120/jacmp.v9i1.2784 |
Ejemplares similares
-
Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database
por: Njeh, Christopher F.
Publicado: (2015) -
Planning quality and delivery efficiency of sMLC delivered IMRT treatment of oropharyngeal cancers evaluated by RTOG H‐0022 dosimetric criteria
por: Zhu, X. Ronald, et al.
Publicado: (2004) -
Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy
por: Wang, Xin, et al.
Publicado: (2016) -
The accuracy of dynamic wedge dose computation in the ADAC Pinnacle RTP system
por: Shao, H., et al.
Publicado: (2004) -
Behavior of the Siemens Virtual Wedge™ following an interruption to beam delivery
por: Richmond, N. D., et al.
Publicado: (2003)